D
(ecimal)
Point ["." / ","]
Identical to Channel A
PWR
CAL
(ibration)
Starts the Power Meter calibration routine. See the appropriate
chapter on how to perform the calibration.
Calibration
By performing the calibration routine using precise standards,
the accuracy of the time base and the Power Meter can be
increased further.
Please let the counter warm up for at least
30 min before performing the calibration!
Time Base (Channel A/B)
The calibration of the time base can be done in two ways:
manually by entering an offset value in ppm (parts per million)
or automatically with a precise 10 MHz reference. After the
calibration routine is activated, "CAL. SELECT" is shown.
40