10
Pool Water Connection
The pool water to be measured must be connected to ASIN AQUA Oxygen.
Connect the measured water inlet downstream of the pump, upstream of the
fi lter and ZPM (coagulation mixer).
Place the water valve fi tting to the T-joint, blinded by DN50 G1/4” threaded plug
#12134.
•
The connecting fi tting is installed in G1/4” thread.
Aseko provides the unique Speedfi t connecting fi tting for measured water
connection to your ASIN AQUA Oxygen. To connect the interconnecting
tube, push it into the Speedfi t connector and then lock it by pulling back. To
disconnect the interconnecting tube, push and hold a circular collet and pull the
interconnecting tube out.
WARNING:
To treat PE tube ends, use a sharp knife (part of delivery). The use
of scissors or tongs deforms the tube end and leads to leaks.
WARNING:
Only hand tighten. Do not use tongs or any other tools.
1.
To ensure tightness of joints, cut the tube at 90 ° angle.
•
Use the sharp knife to cut plastic. The cut must be clean and smooth.
2.
Connect the water inlet to the measured water fi lter connection and the
water outlet to the probe housing connection. Make sure that measured
water pressure does not exceed 1.5 bar.
3.
The water outlet can be realized:
1. downstream of fi ltration
2. to non-pressure (overfl ow tank)
3. upstream of fi lter pump
fi tting
G1/4” thread
6.2x4 hose
CONNECTION
LOCKING
DISCONNECTION
Water valve fi tting
Inserting DN50
plug
G1/4” threaded
T-joint
6.2x4 hose
90°
Pull out