8
AEU-350 / AEU-350S
SYSTEM SETUP
The items identified in the
Maintenance and Sterilization
section (page
14) must be sterilized before each use.
1
Lay the unit on its back. Release the latches
A
and open the
case to remove the components stored inside. Confirm that all
of the components shown in
Package Contents
(page 4) are
present.
A
2
Position the unit so that you can see the bottom side of the
bulkhead. Note that the syringe lines
A
are permanently con-
nected to the bulkhead. The AEU-350 features an auxiliary
water port
B
, and the AEU-350S features a scaler which is per-
manently attached at this location
B
.
A
B
3
Connect the footswitch plug
A
to its socket on the bulkhead,
behind the air tank
B
.
The footswitch plug has a notch
C
and can only be
inserted into the socket in one orientation. Hold the plug
against the socket and rotate it until you feel the plug
enter the socket, then turn the locking sleeve clockwise
until it clicks, securing the connection.
A
C
B
4
Connect the saliva ejector line
A
simply by pressing the con-
nector straight into the socket.
A
5
Connect the HVE line
A
simply by pressing the connector
straight into the socket.
A