UNIT OPERATION:

1

.

Handpieces 

(sold separately)

- Connect

the proper highspeed and lowspeed

handpieces to their appropriate connectors.

Lift handpiece from holder to automatically

activate.  Depress foot control for

operation.

2. 3-Way Air/Water Syringe

- Depress the

right button for air operation, and the left

button for water operation. Depressing

both buttons will create a mist. The syringe

features quick-change autoclavable tips:

To remove a tip, press the locking ring

surrounding the tip socket and pull the

used tip straight out of the socket (Fig. 5).

To insert a new tip, press locking ring and

push tip into socket as far as it will go.

Release ring and gently tug on tip before

using to ensure that tip is securely locked

into socket.

Syringe Tip Sterilization:

1)

Remove contaminated syringe tip.

2)

Remove all visible signs of

contamination before autoclaving.

3)

Autoclave tip at 132° C (270° F) for ten

minutes.

4)

Sterilize between each patient use.

NOTE

: Since only the tips can be

autoclaved, it is recommended that the

air/water syringe be bagged with a

disposable, single-use plastic sleeve

between each patient use.

OPERATION FUNCTIONS:

(See Figures 3 & 4)

1. Handpiece Selectors

- Automatically

activates highspeed or lowspeed

handpieces when lifted from their holders.

2. Pressure Guage

- Provides visual

indication of drive air pressure for the

handpiece in use.

3. Drive Air Pressure Controls

- Adjusts

drive air pressure to the highspeed and

lowspeed handpieces. Turn controls

clockwise (with small slot-head

screwdriver) to decrease pressure;

counterclockwise to increase pressure.

4. Air/Water Syringe

- Delivers air, water, or

a combination mist.  Supplied with

autoclavable tips.

5. Water Flow Controls

- Adjusts water

coolant flow to the highspeed handpieces.

6. Foot Control w/Water On/Off Switch

-

Wet/dry type foot control provides drive air

pressure to the active handpiece.  Apply

light foot pressure to any part of the foot

control cover. 

Toggle Switch

provides

on/off control of water coolant flow to

highspeed handpieces (Fig. 3).

PRESSURE

GUAGE

DRIVE AIR

PRESSURE

CONTROLS

HIGHSPEED

HANDPIECE

SELECTORS 

LOWSPEED

HANDPIECE

SELECTOR

WATER FLOW

CONTROLS

(Underneath

Console)

AIR/WATER

SYRINGE

Figure 4 - Console 

Figure 5 - Tip Removal & Installation

DEPRESS

LOCKING RING