UNIT OPERATION:
1
.
Handpieces
(sold separately)
- Connect
the proper highspeed and lowspeed
handpieces to their appropriate connectors.
Lift handpiece from holder to automatically
activate. Depress foot control for
operation.
2. 3-Way Air/Water Syringe
- Depress the
right button for air operation, and the left
button for water operation. Depressing
both buttons will create a mist. The syringe
features quick-change autoclavable tips:
To remove a tip, press the locking ring
surrounding the tip socket and pull the
used tip straight out of the socket (Fig. 5).
To insert a new tip, press locking ring and
push tip into socket as far as it will go.
Release ring and gently tug on tip before
using to ensure that tip is securely locked
into socket.
Syringe Tip Sterilization:
1)
Remove contaminated syringe tip.
2)
Remove all visible signs of
contamination before autoclaving.
3)
Autoclave tip at 132° C (270° F) for ten
minutes.
4)
Sterilize between each patient use.
NOTE
: Since only the tips can be
autoclaved, it is recommended that the
air/water syringe be bagged with a
disposable, single-use plastic sleeve
between each patient use.
OPERATION FUNCTIONS:
(See Figures 3 & 4)
1. Handpiece Selectors
- Automatically
activates highspeed or lowspeed
handpieces when lifted from their holders.
2. Pressure Guage
- Provides visual
indication of drive air pressure for the
handpiece in use.
3. Drive Air Pressure Controls
- Adjusts
drive air pressure to the highspeed and
lowspeed handpieces. Turn controls
clockwise (with small slot-head
screwdriver) to decrease pressure;
counterclockwise to increase pressure.
4. Air/Water Syringe
- Delivers air, water, or
a combination mist. Supplied with
autoclavable tips.
5. Water Flow Controls
- Adjusts water
coolant flow to the highspeed handpieces.
6. Foot Control w/Water On/Off Switch
-
Wet/dry type foot control provides drive air
pressure to the active handpiece. Apply
light foot pressure to any part of the foot
control cover.
Toggle Switch
provides
on/off control of water coolant flow to
highspeed handpieces (Fig. 3).
PRESSURE
GUAGE
DRIVE AIR
PRESSURE
CONTROLS
HIGHSPEED
HANDPIECE
SELECTORS
LOWSPEED
HANDPIECE
SELECTOR
WATER FLOW
CONTROLS
(Underneath
Console)
AIR/WATER
SYRINGE
Figure 4 - Console
Figure 5 - Tip Removal & Installation
DEPRESS
LOCKING RING