2

SETTING UP THE UNIT:

The ADU-08SX unit may be set up independ-

ently or clamped to a table or counter top.

1. Unpack the ADU-08SX unit, and set on a

table top.

2. Open the unit case and remove the Quick

Grip bar clamp.

3. Position the unit close to the edge of the

table top.  Expand Quick Grip bar clamp,

position over bottom edge of unit case and

under the edge of the table, then ratchet

trigger to compress and secure unit to

table.

4. Fill plastic water bottle with clean water and

screw into top mount beneath the unit

control.

5. Remove Air Vacuum System Assembly

from case and insert male quick disconnect

into air receptacle on the side of the case.

Hang hose end in holder on the front of the

unit.

6. Place the air/water syringe into the proper

holder on the front of the unit.  Place the

disc foot control in a comfortable position

on the floor.

7. Extend 1/4" grey air supply line with quick

disconnect and plug into the AA-74

compressor, or a clean 50-100 PSI

compressed air supply.

OPERATION:

After the unit has been set up, you are ready

for operation as follows:

TA-90D Air/Water Syringe

- Depress the right

button for air operation, and the left button for

water operation. Depressing both buttons will

create a mist. The syringe features quick-

change autoclavable tips: To remove a tip,

press on the locking collar surrounding the tip

socket and pull the used tip straight out of the

socket. To insert a new tip, press locking collar

and push tip into socket as far as it will go.

Release collar and gently tug on tip before

using to ensure that tip is securely locked into

socket.

3

VACUUM HOSE

SYRINGE TIP REMOVAL

DEPRESS

LOCKING COLLAR

Syringe Tip Sterilization:

1)

Remove contaminated syringe tip.

2)

Remove all visible signs of contamination

before autoclaving.

3)

Autoclave tip at 132° C (270° F) for ten minutes.

4)

Sterilize between each patient use.

NOTE

: Since only the tips can be auto-

claved, it is recommended that the air/water

syringe be bagged with a disposable, single-

use plastic sleeve between each patient use.

AA-21SX Air Vacuum System

- When the

unit is connected to a compressed air

source, and the vacuum bottle assembly is

attached to the air receptacle on the side of

the unit case, the Air Vacuum System should

be fully functional.  Insert a standard

disposable vacuum tip into the hose end.

Depress button on hose end to activate

vacuum.  Observe waste bottle while in use

and dispose of waste when bottle is

approximately 3/4 full.
Avoid waste flow-back during use by keeping

the ON/OFF button depressed prior to

entering the oral cavity and until the hose

end has been removed from the patients

mouth.  Clean the entire hose assembly and

waste bottle after each use by evacuating a

cleaner/disinfectant through the system and

thoroughly scrubbing entire assembly.

Summary of Contents for ADU-08SX Sealant Express

Page 1: ...eplace any defective component or product in part or whole Aseptico shall be the sole arbiter of such action In the event of alleged defect under warranty the purchaser is to notify Aseptico s Custome...

Page 2: ...tients and used tips thoroughly cleaned and autoclaved before reuse Air Water Filter The ADU 08SX unit contains a built in air line filter It removes unwanted moisture oil dirt chips and other contami...

Page 3: ...truction Manual 4 MAINTENANCE Master Pressure The master pressure regulator controls the incoming supply pressure It is located behind the vacuum holder underneath the control panel inside the unit Th...

Page 4: ...ress on the locking collar surrounding the tip socket and pull the used tip straight out of the socket To insert a new tip press locking collar and push tip into socket as far as it will go Release co...

Page 5: ...ress on the locking collar surrounding the tip socket and pull the used tip straight out of the socket To insert a new tip press locking collar and push tip into socket as far as it will go Release co...

Page 6: ...truction Manual 4 MAINTENANCE Master Pressure The master pressure regulator controls the incoming supply pressure It is located behind the vacuum holder underneath the control panel inside the unit Th...

Page 7: ...tients and used tips thoroughly cleaned and autoclaved before reuse Air Water Filter The ADU 08SX unit contains a built in air line filter It removes unwanted moisture oil dirt chips and other contami...

Page 8: ...eplace any defective component or product in part or whole Aseptico shall be the sole arbiter of such action In the event of alleged defect under warranty the purchaser is to notify Aseptico s Custome...

Reviews: