TABLE OF CONTENTS:

Package Contents  . . . . . . . . . . . . . . . . . . . .i

Introduction  . . . . . . . . . . . . . . . . . . . . . . . . .1

Setting Up the Unit  . . . . . . . . . . . . . . . . . . .1

Operation . . . . . . . . . . . . . . . . . . . . . . . . . . .3

Maintenance  . . . . . . . . . . . . . . . . . . . . . . . .5

Warranty  . . . . . . . . . . . . . . . . . . .Back Cover

Size:

12.5” x 17.5” x 7.5”

(32 cm x 44 cm x 19 cm)

Weight:

32.6 lbs (14.8 kg)

Power Source:

115 VAC, 60 Hz or

230 VAC, 50 Hz

Current Rating:

6.3A at 115 VAC or

4A at 230 VAC

Fuse:

115V Operation:

6.3A, 125V

Medium Time Delay

230V Operation:

4A, 250V Slo-

Blo Type

Air Source:

1/2 H.P. Oil-Free 

Compressor/Vacuum Pump

Water Supply Tank: 900 mL (0.95 Quart)

Waste Tank:

946 mL (1 Quart)

Vacuum:

1.35 CFM @ 4.0 In. Hg.

Environmental Conditions:

Intended for indoor use in an office environment of

10°C - 40°C, 10% - 90% humidity, non-condensing.

SPECIFICATIONS:

#TA-1

TA-1 Syringe Tip

#AA-12A

Handpiece Holder

Extension Arm

#AA-12CSB

Case Stand-Off

Bracket

#AA-12HVE

HVE Tip Adapter

#AA-26SS

Stainless Steel 

Dry Oral Cup

Available Accessories:

PACKAGE CONTENTS

ADU-12DCEB System Case/Control Box

Highspeed and Lowspeed Handpiece Controls

TA-90D 3-Way Air/Water Syringe with Tip

NWS-8 Water Supply Tank

730068-08 Waste Container

Electric Foot Switch

Saliva Ejector Vacuum System

Remote AC Power Cord

AA-12CSB Case Stand-Off Bracket

Two (2) Extra Fuses

Shoulder Strap

Manual

P.O. Box 1548 • Woodinville, WA 98072

8333 216

th

Street S.E. • Woodinville, WA 98072

International (425) 487-3157 • Toll Free (800) 426-5913

www.aseptico.com • info@aseptico.com

i

To prevent injury to people and damage to

property, please heed relevant warnings and

remarks. They are marked as follows:

WARNING:

Serious injury or death may result if

ignored.

CAUTION:

Damage to property or the

environment may result if ignored.

NOTE:

Important additional information

and hints.

WARNING: This device has been tested and found to comply with

the emissions requirements of IEC 60601-1-2:2001-09.  These

requirements provide reasonable protection against harmful

electromagnetic interference in a typical medical installation.

However, high levels of radio-frequency (RF) emissions from

electrical devices, such as cellular phones, may disrupt the

performance of this device.  To mitigate disruptive electromagnetic

interference, position this device away from RF transmitters and

other sources of electromagnetic energy.

WARNING:

Device not suitable for use in the presence of

flammable gasses or gasseous 

oxidizers.

Summary of Contents for ADU-12DCEB Task Force Deluxe

Page 1: ...ADU 12DCEB Task Force Deluxe Portable Pneumatic Dental System OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Page 2: ...ot Switch Saliva Ejector Vacuum System Remote AC Power Cord AA 12CSB Case Stand Off Bracket Two 2 Extra Fuses Shoulder Strap Manual P O Box 1548 Woodinville WA 98072 8333 216th Street S E Woodinville...

Page 3: ...at stable surface 2 Unlatch and open the case Install AA 12CSB Bracket to hold up case lid Figure 1 3 Verify that the voltage selector insert is configured properly by checking the voltage indicator o...

Page 4: ...r the country of usage and carries the proper certifications Carefully reinstall fuses into fuse holder and slide holder back into selector module Figure 6 Apply pressure until locking tab snaps back...

Page 5: ...sing and depressing the foot switch within two seconds may cause the compressor vacuum pump to stall and blow the fuses when the handpiece on off toggle switch is in the OFF position To prevent this w...

Page 6: ...er system Then unscrew the cap from tank and fill tank with water Screw cap back onto tank Sterilization of tank is recommended between patient use SALIVA EJECTOR VACUUM SYSTEM The Saliva Ejector is p...

Page 7: ...low bleach solution to stand in lines for 10 minutes Remove water supply tank and discard bleach Flush water supply tank and all lines thoroughly with clean water Air purge and leave lines dry until n...

Page 8: ...co s Customer Service Department promptly Customer Service will provide instructions usually directing that the product be returned for service Shipment to Aseptico and the cost thereof is always the...

Reviews: