background image

2

SALIVA EJECTOR VACUUM

WATER COOLANT ON/OFF

TOGGLE SWITCH

WATER COOLANT

FLOW CONTROL

HANDPIECE SELECTION

TOGGLE SWITCH

SETTING UP THE UNIT

..Cont’d

IMPORTANT: For 110V-120V use, check that the

fuse in use is a 6.3A medium time delay; for

220V-250V use, check that the fuse is a 4A Slo-

Blo type. User may have to change fuses

before use.

For 220V-250V use, confirm that the power cord

is correct for the country of usage and carries the

proper certifications.

Carefully reinstall fuses into fuse holder and slide

holder back into selector module 

(Figure 6)

. Apply

pressure until locking tab snaps back into

position, producing a distinct clicking sound. 

WASTE

CONTAINER

CAVITRON WATER

OUTLET

HANDPIECE ON/OFF

TOGGLE SWITCH

TANK PRESSURE TOG-

GLE SWITCH

FOOT

SWITCH

WATER

SUPPLY

TANK

Fig. 5

Fig. 3

Fig. 4

Fig. 6

FUSES LOCATED

IN VOLTAGE

SELECTOR

MODULE

3-WAY

AIR/WATER

SYRINGE

HANDPIECE

CONTROLS

HOLDER BAR

AA-12CSB

BRACKET

VOLTAGE

SELECTOR

MODULE

& FUSE

HOLDER

LOCATION

SLIDING

PANEL

Fig. 7 (Rear View)

Summary of Contents for ADU-12DCEB Task Force Deluxe

Page 1: ...ADU 12DCEB Task Force Deluxe Portable Pneumatic Dental System OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Page 2: ...ot Switch Saliva Ejector Vacuum System Remote AC Power Cord AA 12CSB Case Stand Off Bracket Two 2 Extra Fuses Shoulder Strap Manual P O Box 1548 Woodinville WA 98072 8333 216th Street S E Woodinville...

Page 3: ...at stable surface 2 Unlatch and open the case Install AA 12CSB Bracket to hold up case lid Figure 1 3 Verify that the voltage selector insert is configured properly by checking the voltage indicator o...

Page 4: ...r the country of usage and carries the proper certifications Carefully reinstall fuses into fuse holder and slide holder back into selector module Figure 6 Apply pressure until locking tab snaps back...

Page 5: ...sing and depressing the foot switch within two seconds may cause the compressor vacuum pump to stall and blow the fuses when the handpiece on off toggle switch is in the OFF position To prevent this w...

Page 6: ...er system Then unscrew the cap from tank and fill tank with water Screw cap back onto tank Sterilization of tank is recommended between patient use SALIVA EJECTOR VACUUM SYSTEM The Saliva Ejector is p...

Page 7: ...low bleach solution to stand in lines for 10 minutes Remove water supply tank and discard bleach Flush water supply tank and all lines thoroughly with clean water Air purge and leave lines dry until n...

Page 8: ...co s Customer Service Department promptly Customer Service will provide instructions usually directing that the product be returned for service Shipment to Aseptico and the cost thereof is always the...