OPERATION:

FOOT SWITCH

All Task Force Deluxe system components are

activated by the on/off foot switch. The foot switch

enables the operation of the handpieces, saliva

ejector and the 3-way air/water syringe.

COMPRESSED AIR SUPPLY

The ADU-12DCEB contains a 1/2 HP compressor/

vacuum pump that powers all functions. The

compressed air supply and vacuum are available

only when the foot switch is depressed.

SETTING UP THE UNIT

..Cont’d

4. Place the syringe, highspeed handpiece and

lowspeed handpiece in their proper holders on the

holder bar.

5. Place the foot switch on the floor.
6. Remove the cap from the white plastic water

supply tank and fill with clean water. Reinstall the

cap and place water supply tank into retainer strap

inside Task Force Deluxe case.

7. Attach the saliva ejector hose onto the top of the

waste container by pushing in and rotating 1/4 turn

(Figure 8)

. Verify that the vacuum line and waste

container are properly installed. Hang the saliva

ejector hose in its holder on the holder bar.

Fig. 8

TO INSTALL SALIVA EJECTOR HOSE,

TURN CONNECTOR 1/4 TURN CLOCKWISE

CAUTION:

Releasing and depressing the foot switch within

two  seconds  may  cause  the  compressor/

vacuum pump to stall and blow the fuses when

the handpiece on/off toggle switch is in the ‘OFF’

position.  To  prevent  this,  wait  two  or  more

seconds  after  releasing  footswitch  before

depressing it again.

VACUUM LINE

WITH QUICK

DISCONNECT

THREE-WAY AIR/WATER SYRINGE

The ADU-12DCEB Task Force Deluxe features a

TA-90D three-way air/water syringe with quick

change autoclavable tips. To remove a tip, press

on the locking collar surrounding the tip socket and

pull the used tip straight out of the socket (see Fig.

9). To insert a new tip, press locking collar and

push tip into socket as far as it will go. Release

collar and gently tug on tip before using to ensure

that tip is securely locked into socket.

Syringe Tip Sterilization:

1)

Remove contaminated syringe tip.

2)

Remove all visible signs of 

contamination before autoclaving.

3)

Autoclave tip at 132° C (270° F) for ten

minutes.

4)

Sterilize between each patient use.

NOTE

: Since only the tips can be autoclaved, it is

recommended that the Air/Water Syringe be bagged

with a disposable, single-use plastic sleeve between

each patient use.

Avoid damage to the syringe by removing the tip when

closing the ADU-12DCEB case for transportation.

NOTE

FIG. 9

DEPRESS

LOCKING COLLAR

Summary of Contents for ADU-12DCEB Task Force Deluxe

Page 1: ...ADU 12DCEB Task Force Deluxe Portable Pneumatic Dental System OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Page 2: ...ot Switch Saliva Ejector Vacuum System Remote AC Power Cord AA 12CSB Case Stand Off Bracket Two 2 Extra Fuses Shoulder Strap Manual P O Box 1548 Woodinville WA 98072 8333 216th Street S E Woodinville...

Page 3: ...at stable surface 2 Unlatch and open the case Install AA 12CSB Bracket to hold up case lid Figure 1 3 Verify that the voltage selector insert is configured properly by checking the voltage indicator o...

Page 4: ...r the country of usage and carries the proper certifications Carefully reinstall fuses into fuse holder and slide holder back into selector module Figure 6 Apply pressure until locking tab snaps back...

Page 5: ...sing and depressing the foot switch within two seconds may cause the compressor vacuum pump to stall and blow the fuses when the handpiece on off toggle switch is in the OFF position To prevent this w...

Page 6: ...er system Then unscrew the cap from tank and fill tank with water Screw cap back onto tank Sterilization of tank is recommended between patient use SALIVA EJECTOR VACUUM SYSTEM The Saliva Ejector is p...

Page 7: ...low bleach solution to stand in lines for 10 minutes Remove water supply tank and discard bleach Flush water supply tank and all lines thoroughly with clean water Air purge and leave lines dry until n...

Page 8: ...co s Customer Service Department promptly Customer Service will provide instructions usually directing that the product be returned for service Shipment to Aseptico and the cost thereof is always the...