OPERATION:
FOOT SWITCH
All Task Force Deluxe system components are
activated by the on/off foot switch. The foot switch
enables the operation of the handpieces, saliva
ejector and the 3-way air/water syringe.
COMPRESSED AIR SUPPLY
The ADU-12DCEB contains a 1/2 HP compressor/
vacuum pump that powers all functions. The
compressed air supply and vacuum are available
only when the foot switch is depressed.
SETTING UP THE UNIT
..Cont’d
4. Place the syringe, highspeed handpiece and
lowspeed handpiece in their proper holders on the
holder bar.
5. Place the foot switch on the floor.
6. Remove the cap from the white plastic water
supply tank and fill with clean water. Reinstall the
cap and place water supply tank into retainer strap
inside Task Force Deluxe case.
7. Attach the saliva ejector hose onto the top of the
waste container by pushing in and rotating 1/4 turn
(Figure 8)
. Verify that the vacuum line and waste
container are properly installed. Hang the saliva
ejector hose in its holder on the holder bar.
Fig. 8
TO INSTALL SALIVA EJECTOR HOSE,
TURN CONNECTOR 1/4 TURN CLOCKWISE
CAUTION:
Releasing and depressing the foot switch within
two seconds may cause the compressor/
vacuum pump to stall and blow the fuses when
the handpiece on/off toggle switch is in the ‘OFF’
position. To prevent this, wait two or more
seconds after releasing footswitch before
depressing it again.
VACUUM LINE
WITH QUICK
DISCONNECT
THREE-WAY AIR/WATER SYRINGE
The ADU-12DCEB Task Force Deluxe features a
TA-90D three-way air/water syringe with quick
change autoclavable tips. To remove a tip, press
on the locking collar surrounding the tip socket and
pull the used tip straight out of the socket (see Fig.
9). To insert a new tip, press locking collar and
push tip into socket as far as it will go. Release
collar and gently tug on tip before using to ensure
that tip is securely locked into socket.
Syringe Tip Sterilization:
1)
Remove contaminated syringe tip.
2)
Remove all visible signs of
contamination before autoclaving.
3)
Autoclave tip at 132° C (270° F) for ten
minutes.
4)
Sterilize between each patient use.
NOTE
: Since only the tips can be autoclaved, it is
recommended that the Air/Water Syringe be bagged
with a disposable, single-use plastic sleeve between
each patient use.
Avoid damage to the syringe by removing the tip when
closing the ADU-12DCEB case for transportation.
NOTE
FIG. 9
DEPRESS
LOCKING COLLAR