4

HANDPIECE CONTROLS

The Aseptico ADU-12DCEB includes a manual

handpiece control system. The system includes

one highspeed control with water coolant and one

lowspeed control. The handpiece selection switch

allows the user to select highspeed or lowspeed

operation, and the handpiece on/off toggle switch

turns the selected handpiece on or off.

TANK PRESSURE TOGGLE SWITCH

Turns the pressure to the water supply tank on or

off. Turn switch off before removing water supply

tank.

WATER COOLANT ON/OFF TOGGLE SWITCH

Turns water coolant to highspeed handpiece on or

off. 

WATER COOLANT FLOW CONTROL

Adjusts the flow of water coolant to the highspeed

handpiece control from off to full spray when the

water coolant on/off toggle switch is in the “ON”

position.

CAVITRON WATER OUTLET

Provides water via standard 1/4” quick disconnect

to independent Cavitron or scaler units requiring a

water supply.

WATER COOLANT

FLOW CONTROL

WATER COOLANT

ON/OFF

TOGGLE SWITCH

PRESSURIZED WATER SUPPLY SYSTEM

The Aseptico ADU-12DCEB incorporates a self

contained pressurized water system. This consists

of a 900 mL white plastic tank dispensing water

through the 3-way syringe and highspeed

handpiece control. Before removing the tank for

refilling, it must be depressurized. Turn off the tank

pressure toggle switch to release pressure in the

water system. Then unscrew the cap from tank

and fill tank with water. Screw cap back onto tank.

Sterilization of tank is recommended between

patient use.

SALIVA EJECTOR VACUUM SYSTEM

The Saliva Ejector is powered by the

compressor/vacuum pump and is a self contained

oral evacuation system. It is supplied with a plastic

waste container and a saliva ejector hose.
To operate, attach the saliva ejector hose to the

top of the waste container and insert a standard

saliva ejector tip. To activate the saliva ejector,

depress the foot switch.
Waste from the vacuum system will collect in the

clear plastic waste container. When the waste

container is over 1/2 full, it must be emptied to

prevent a back-up in the system. After each use,

the vacuum system should be cleaned by running

clean water or vacuum system cleanser through

the saliva ejector hose and by thoroughly cleaning

the waste container. 

CAUTION: 

Use only NON-

foaming cleansers in the vaccum lines. Use a new

saliva ejector tip for each patient. 

HANDPIECE

SELECTION

SWITCH

HANDPIECE

ON/OFF

TOGGLE SWITCH

CAVITRON

WATER OUTLET

TANK PRESSURE

TOGGLE SWITCH

THE WATER SUPPLY TANK IS

PRESSURIZED DURING

OPERATION. DEPRESSURIZE

WATER TANK BY TURNING OFF

TANK PRESSURE TOGGLE

SWITCH BEFORE REMOVAL.

WARNING:

The waste container has a ball float shutoff to

prevent accidental overflow. If the waste

container becomes full and the ball float

activates, the vacuum motor may not restart,

and could potentially blow the fuse.

Immediately empty waste container to

continue operation.

NOTE:

Summary of Contents for ADU-12DCEB Task Force Deluxe

Page 1: ...ADU 12DCEB Task Force Deluxe Portable Pneumatic Dental System OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Page 2: ...ot Switch Saliva Ejector Vacuum System Remote AC Power Cord AA 12CSB Case Stand Off Bracket Two 2 Extra Fuses Shoulder Strap Manual P O Box 1548 Woodinville WA 98072 8333 216th Street S E Woodinville...

Page 3: ...at stable surface 2 Unlatch and open the case Install AA 12CSB Bracket to hold up case lid Figure 1 3 Verify that the voltage selector insert is configured properly by checking the voltage indicator o...

Page 4: ...r the country of usage and carries the proper certifications Carefully reinstall fuses into fuse holder and slide holder back into selector module Figure 6 Apply pressure until locking tab snaps back...

Page 5: ...sing and depressing the foot switch within two seconds may cause the compressor vacuum pump to stall and blow the fuses when the handpiece on off toggle switch is in the OFF position To prevent this w...

Page 6: ...er system Then unscrew the cap from tank and fill tank with water Screw cap back onto tank Sterilization of tank is recommended between patient use SALIVA EJECTOR VACUUM SYSTEM The Saliva Ejector is p...

Page 7: ...low bleach solution to stand in lines for 10 minutes Remove water supply tank and discard bleach Flush water supply tank and all lines thoroughly with clean water Air purge and leave lines dry until n...

Page 8: ...co s Customer Service Department promptly Customer Service will provide instructions usually directing that the product be returned for service Shipment to Aseptico and the cost thereof is always the...

Reviews: