MAINTENANCE:

Because of its simple design, the Aseptico ADU-12DCEB Task Force Deluxe requires very little maintenance. Any

maintenance that is needed can be performed in minutes.

BLEEDING THE SYSTEM

If the unit will not be used for an extended period of time, or the unit might be subjected to freezing conditions, you

should bleed the system. Simply empty the contents of the water supply tank and install the cap back onto the tank.

Operate the air/water syringe and highspeed handpiece with water coolant 'on' until just air comes through the water

lines. Pack unit and store as normal.

HANDPIECE FLUSH

Flush the handpieces for about 5 seconds after every patient, and about 20 seconds at the beginning of each day. 

GENERAL CLEANING

The external surfaces of the case should be cleaned using a soft cloth moistened with a mild detergent solution. Any

external surfaces of the unit that are contacted during use should be wiped down with a soft cloth moistened with a

disinfectant at the beginning of each day and between each patient use.
As described in the vacuum section, the saliva ejector vacuum system should be thoroughly cleaned after each use.

WATER LINES

Disinfect the water lines weekly.  Prepare a 1:10 bleach solution (1 part household bleach to 10 parts water). Remove

water reservoir and discard residual water. Replace empty water supply tank and air purge all waterlines. Fill water

supply tank with bleach solution. Run bleach solution through all lines. Allow bleach solution to stand in lines for 10

minutes. Remove water supply tank and discard bleach. 

Flush water supply tank and all lines thoroughly

with clean water

. Air purge and leave lines dry until next clinical use.

Problem:

Unit will not start

Unit starts but blows fuse

No water pressure

Insufficient vacuum

Insufficient handpiece operation

Correction:

Check console to power connection.

Depress foot switch for operation.

Check that voltage selector module is set for source voltage.

Check fuse.  If blown, replace with 6.3A/125V medium time delay fuse for 110V-120V

operation, or a 4A/250V slo-blo fuse for 220V-250V operation.

Check that voltage selector module is set for source voltage.

Check vacuum waste container. If full, immediately empty to continue operation.

Check vacuum hose assembly for blockage. NEVER install a vacuum valve with a

shut-off control.

Depressurize water supply tank by turning off tank pressure toggle switch.

Check that fuse is correct for source voltage.

Check water supply tank for water level and verify that cap is tight.

Turn tank pressure toggle switch “ON”.

Depress foot switch for 5+ sec. to pressurize supply tank.

Turn water coolant flow knob counterclockwise to open.

Check vacuum waste container. If full, immediately empty to continue operation.

Check vacuum hose assembly for blockage. NEVER install a vacuum valve with a

shut-off control.

Check that waste container lid is properly seated and tightly secured.

Check that handpiece ON/OFF toggle is “ON”.

Check that handpiece “SELECT” toggle is set to desired handpiece.

Check that handpiece tubing is untangled and not crimped.

Check handpiece connection for missing gasket.

TROUBLESHOOTING:

Summary of Contents for ADU-12DCEB Task Force Deluxe

Page 1: ...ADU 12DCEB Task Force Deluxe Portable Pneumatic Dental System OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Page 2: ...ot Switch Saliva Ejector Vacuum System Remote AC Power Cord AA 12CSB Case Stand Off Bracket Two 2 Extra Fuses Shoulder Strap Manual P O Box 1548 Woodinville WA 98072 8333 216th Street S E Woodinville...

Page 3: ...at stable surface 2 Unlatch and open the case Install AA 12CSB Bracket to hold up case lid Figure 1 3 Verify that the voltage selector insert is configured properly by checking the voltage indicator o...

Page 4: ...r the country of usage and carries the proper certifications Carefully reinstall fuses into fuse holder and slide holder back into selector module Figure 6 Apply pressure until locking tab snaps back...

Page 5: ...sing and depressing the foot switch within two seconds may cause the compressor vacuum pump to stall and blow the fuses when the handpiece on off toggle switch is in the OFF position To prevent this w...

Page 6: ...er system Then unscrew the cap from tank and fill tank with water Screw cap back onto tank Sterilization of tank is recommended between patient use SALIVA EJECTOR VACUUM SYSTEM The Saliva Ejector is p...

Page 7: ...low bleach solution to stand in lines for 10 minutes Remove water supply tank and discard bleach Flush water supply tank and all lines thoroughly with clean water Air purge and leave lines dry until n...

Page 8: ...co s Customer Service Department promptly Customer Service will provide instructions usually directing that the product be returned for service Shipment to Aseptico and the cost thereof is always the...

Reviews: