background image

9.

Problem:

Unit will not start

Unit starts but blows fuse

No water coming out

Insufficient vacuum

Insufficient handpiece operation

Compressor makes a chugging

sound and does not start up

properly

Correction:

Check system’s power connection.

Turn Run Mode Switch to “Continuous” or “Foot Switch” Mode and begin operation.

Check that voltage selector module is set for source voltage.

Check fuse. If blown, replace with 6.3A/125V medium time delay fuse for 120V

operation, or a 4A/250V slo-blo fuse for 230V operation.

Check that voltage selector module is set for source voltage.

Check vacuum waste container. If full, immediately empty to continue operation.

Check vacuum hose assembly for blockage. 

Check that fuse is correct for source voltage.

Check water supply tank for water level and verify that cap is tight.

Depress foot switch for 5+ sec. to pressurize supply tank.

Ensure the handpiece air control valve is not opened all the way.

Check vacuum waste container. If full, immediately empty to continue operation.

Check vacuum hose assembly for blockage. 

Check that waste container lid is properly seated and tightly secured.

Check that handpiece ON/OFF toggle is “ON”.

Check that handpiece coolant ON/OFF toggle is “ON”.

Check that air control valve is turned up adequately

Check that handpiece tubing is untangled and not crimped.

Check handpiece connection for missing gasket.

Check for sufficient power source and avoid long light-grade extension cords

(Refer to Figure 8 Extension Cord Recommendations - below)

TROUBLESHOOTING:

EXTENSION CORD RECOMMENDATIONS:

Use only a 3-wire extension cord that has

a 3-blade grounding plug. Connect

extension cord plug to a matching 3-slot

receptacle. Ensure that the extension

cord is in good condition and that its gage

wire is the correct size wire to carry the

current this product will draw. An

undersized cord is a potential fire hazard

and will cause a drop in line voltage,

resulting in loss of power and causing the

product to overheat. Figure 9 indicates

the correct size cord for the length

required and the ampere rating listed on

the product nameplate. 

If in doubt, use

the next heavier gage cord. The

smaller the gage number, the heavier

the wire gage.

FIG. 9

Summary of Contents for ADU-17X Express II

Page 1: ...ADU 17X Express II Portable Dental System OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Page 2: ...to property please heed relevant warnings and remarks They are marked as follows WARNING Serious injury or death may result if ignored CAUTION Damage to property or the environment may result if ignor...

Page 3: ...provide many years of reliable service Please read the instructions provided in this manual to receive the best and longest service from your Aseptico equipment Separate manuals may be provided to cov...

Page 4: ...g WARNING Use for intended purposes only Failure to observe the operating instructions may result in the patient or user suffering serious injury or the the product being damaged possibly beyond repai...

Page 5: ...HVE VOLTAGE SELECTOR MODULE WASTE TANK WASTE TANK TIE DOWN WATER SUPPLY BOTTLE LOW VOLUME VACUUM SALIVA EJECTOR FOOT SWITCH RUN MODE SWITCH CASE HANDLE RESERVOIR CONNECTOR COMPRESSOR AND FAN HOUSING...

Page 6: ...ing releases Figure 2b Carefully remove fuse housing to reveal Voltage Selector Insert Figure 2c Pull insert out of housing turn it around and reinsert into housing with proper voltage indicator facin...

Page 7: ...Ejector hose to the black connector located on the front of the Tank Note When attaching the Saliva Ejector hose to the top of the Waste Tank push in and rotate 1 4 turn see Figure 5 7 Secure Waste T...

Page 8: ...nd the ball float activates the vacuum motor may not restart and could potentially blow the fuse Immediately empty waste container to continue operation NOTE OPERATION FUNCTIONS 1 FOOT SWITCH RUN MODE...

Page 9: ...ng 6 Adjust the vacuum pressure by moving the valves on both the HVE and saliva ejector vacuum heads 7 Empty the Vacuum Waste Tank when it is 3 4 full After you have set up the Express II unit and hav...

Page 10: ...for 10 minutes Remove water supply tank and discard bleach Flush water supply tank and all lines thoroughly with clean water Air purge and leave lines dry until next clinical use VACUUM SYSTEM The HV...

Page 11: ...operation Check vacuum hose assembly for blockage Check that waste container lid is properly seated and tightly secured Check that handpiece ON OFF toggle is ON Check that handpiece coolant ON OFF tog...

Page 12: ...ce travel pouch with water bottle on top of coiled hoses 6 Coil foot switch cable and place in pouch with foot switch Coil motor drive air line and place in pouch Place power cord and fuses in pouch C...

Page 13: ...pment must not be disposed as unsorted municipal waste and must be collected separately Disposing of your old product Each Member State in the European Union EU requires that the waste from electrical...

Page 14: ...o 60 C 4 to 140 F Relative Humidity 10 to 90 non condensing Altitude 0 to 3048 meters 0 to 10 000 feet Unit Duty Cycle Continuous This device has been tested and found to comply with the emissions req...

Page 15: ...ter of such action In the event of alleged defect under warranty the purchaser is to notify Aseptico s Customer Service Department promptly Customer Service will provide instructions usually directing...

Page 16: ...P N 420627 Rev L ECO 14117 03 2017 PRINTED IN THE U S A P O Box 1548 Woodinville WA 98072 8333 216th Street SE Woodinville WA 98072 425 487 3157 800 426 5913 www aseptico com info aseptico com...

Reviews: