background image

4.

120V

Indicator

Fig. 2a

240V

Indicator

Voltage Selector Module

Fig. 2d

Fig. 2b

Fig. 2c

Fig. 2e

1. Unpack the ADU-17X Case.
2. Verify that the Voltage Selector Insert is

configured properly by checking the voltage

indicator on the Voltage Selector Module 

(Figure

2a)

. The Voltage Module is located on the lower

left of the Motor Cover. The Voltage Selector

Insert must be installed in the “120” position for

120V 60Hz operation or in the “240” position for

230V 50Hz operation. 

To switch the voltage configuration, disconnect

power cord and insert a small flat screwdriver

under the locking tab in the front of Selector

Module. Gently depress tab until the fuse

housing releases 

(Figure 2b)

. Carefully remove

fuse housing to reveal Voltage Selector Insert

(Figure 2c).

Pull insert out of housing, turn it

around and reinsert into housing with proper

voltage indicator facing out 

(Figure 2d)

.   

IMPORTANT: For 120V use, check that the

Fuse in use is a 6.3A medium time delay; for

230V use, check that the Fuse is a 4A Slo-Blo

type. User may have to change fuses before

use.

For 230V use, confirm that the power cord is

correct for the country of usage and carries the

proper certifications.

Carefully reinstall fuses into Fuse Holder and

slide Holder back into Selector Module 

(Figure

1e)

. Apply pressure until locking tab snaps back

into position, producing a distinct clicking sound. 

FUSES LOCATED

IN VOLTAGE

SELECTOR

MODULE

WARNING:

ALWAYS DISCONNECT POWER TO

THE SYSTEM BEFORE OPENING

VOLTAGE SELECTOR MODULE

SETTING UP THE UNIT  

Cont’d

:

Summary of Contents for ADU-17X Express II

Page 1: ...ADU 17X Express II Portable Dental System OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Page 2: ...to property please heed relevant warnings and remarks They are marked as follows WARNING Serious injury or death may result if ignored CAUTION Damage to property or the environment may result if ignor...

Page 3: ...provide many years of reliable service Please read the instructions provided in this manual to receive the best and longest service from your Aseptico equipment Separate manuals may be provided to cov...

Page 4: ...g WARNING Use for intended purposes only Failure to observe the operating instructions may result in the patient or user suffering serious injury or the the product being damaged possibly beyond repai...

Page 5: ...HVE VOLTAGE SELECTOR MODULE WASTE TANK WASTE TANK TIE DOWN WATER SUPPLY BOTTLE LOW VOLUME VACUUM SALIVA EJECTOR FOOT SWITCH RUN MODE SWITCH CASE HANDLE RESERVOIR CONNECTOR COMPRESSOR AND FAN HOUSING...

Page 6: ...ing releases Figure 2b Carefully remove fuse housing to reveal Voltage Selector Insert Figure 2c Pull insert out of housing turn it around and reinsert into housing with proper voltage indicator facin...

Page 7: ...Ejector hose to the black connector located on the front of the Tank Note When attaching the Saliva Ejector hose to the top of the Waste Tank push in and rotate 1 4 turn see Figure 5 7 Secure Waste T...

Page 8: ...nd the ball float activates the vacuum motor may not restart and could potentially blow the fuse Immediately empty waste container to continue operation NOTE OPERATION FUNCTIONS 1 FOOT SWITCH RUN MODE...

Page 9: ...ng 6 Adjust the vacuum pressure by moving the valves on both the HVE and saliva ejector vacuum heads 7 Empty the Vacuum Waste Tank when it is 3 4 full After you have set up the Express II unit and hav...

Page 10: ...for 10 minutes Remove water supply tank and discard bleach Flush water supply tank and all lines thoroughly with clean water Air purge and leave lines dry until next clinical use VACUUM SYSTEM The HV...

Page 11: ...operation Check vacuum hose assembly for blockage Check that waste container lid is properly seated and tightly secured Check that handpiece ON OFF toggle is ON Check that handpiece coolant ON OFF tog...

Page 12: ...ce travel pouch with water bottle on top of coiled hoses 6 Coil foot switch cable and place in pouch with foot switch Coil motor drive air line and place in pouch Place power cord and fuses in pouch C...

Page 13: ...pment must not be disposed as unsorted municipal waste and must be collected separately Disposing of your old product Each Member State in the European Union EU requires that the waste from electrical...

Page 14: ...o 60 C 4 to 140 F Relative Humidity 10 to 90 non condensing Altitude 0 to 3048 meters 0 to 10 000 feet Unit Duty Cycle Continuous This device has been tested and found to comply with the emissions req...

Page 15: ...ter of such action In the event of alleged defect under warranty the purchaser is to notify Aseptico s Customer Service Department promptly Customer Service will provide instructions usually directing...

Page 16: ...P N 420627 Rev L ECO 14117 03 2017 PRINTED IN THE U S A P O Box 1548 Woodinville WA 98072 8333 216th Street SE Woodinville WA 98072 425 487 3157 800 426 5913 www aseptico com info aseptico com...

Reviews: