11
AEU-1070-70V / AEU-1070
1 .
Brush off debris from the motor and cord.
2 .
Thoroughly clean the motor and cord with a moist cloth or
towel to remove any remaining signs of debris. Avoid pulling
the cord away from the motor. Always wipe towards the motor.
3 .
Install the provided autoclaving plug
A
onto the end of the
motor cord connector.
A
4 .
Sterilize the motor and cord assembly using one of the
STERILIZATION METHODS listed below (A, B or C).
Loosely coil the motor cord when autoclaving. Avoid
sharply bending the cord.
STERILIZATION METHODS:
Wrapped Sterilization
— Place in an appropriately sized steriliza-
tion pouch and seal.
A . Standard autoclaving
(Gravity displacement
method)
Time
: 15 min
Temperature
: 132° C (270° F)
Dry time
: 30 minutes
B . Pre-vacuum
(Dynamic air removal method)
Time
: 4 minutes
Temperature
: 132° C (270° F)
Dry time
: 40 minutes
Flash
Sterilization
— For immediate use only.
C . Unwrapped standard autoclaving
(Gravity
displacement method)
Time
: 10 minutes
Temperature
: 132° C (270° F)
Dry time
: No dry time is required for flash
sterilization.
Silicone Water Lines
1 .
Flush the tubing with clean water for 30 seconds after each
use. Do not use any disinfectants on the tubing.
2 .
Sterilize the tubing:
Wrapped Sterilization —
Place in an appropriately
sized sterilization pouch and seal.
Standard autoclaving
(Gravity displacement method)
Time
: 30 min
Temperature
: 132° C (270° F)
Dry time
: 30 minutes
Console and Foot Controller
1 .
Clean the exterior of the console and foot controller by wiping
with a soft cloth moistened with mild detergent or 1:10 bleach
solution (1-part bleach to 9-parts water). Use of solvents or
other solutions may cause damage and void the warranty.