AEU-1070-70V / AEU-1070

CONTROL PANEL

This section describes the functions of the touch-screen.

Home Screen

The home screen displays the name of the active preset   and the 

current settings for each of the adjustable parameters 

:

 

PRESET   —

 This segment of the screen displays the active 

preset name. Press here   to select presets sequentially. 

Below is a table of the factory settings for each preset.

PRESET

DIRECTION SPEED

RATIO

TORQUE

WATER

PRST-1

FORWARD

1200

20:1

50

70

PRST-2

FORWARD

800

20:1

50

70

PRST-3

FORWARD

15

20:1

35

0

PRST-4

FORWARD

15

20:1

50

0

PRST-5

REVERSE

25

20:1

50

0

PRST-6

FORWARD

20

20:1

30

0

PRST-7

FORWARD

21

20:1

40

0

PRST-8

FORWARD

22

20:1

50

0

 

MOTOR DIRECTION  —

 This segment of the screen displays 

the current motor rotation direction. Press here   to alternate 

between forward (F) and reverse (R).

 

SPEED / RATIO   —

 This segment of the screen displays both 

the current speed 

and

 ratio settings. The ratio setting 

(20:1 

or 1:1)

 is indicated by the format 

(whole number or scientific 

notation, respectively)

 in which the speed setting is displayed: 

 

RATIO INDICATOR

: The speed setting is displayed 

as a whole number (e.g. 2000) to indicate when the 

20:1 ratio

 is currently selected. The speed setting is 

displayed in scientific notation (e.g.  40E3) to indicate 

when the 

1:1  ratio

 is currently selected. Replace “E3” 

with “000” to calculate the speed of a 1:1  handpiece 

(e.g. 40E3 = 40,000 RPM).

Press here   to open the speed/ratio  adjustment screen:

• SPEED SETTING:

 

Press the 

 and 

 buttons to adjust the 

speed setting in the following ranges, according to the 

current ratio setting: 

RATIO SETTING

SPEED RANGE

SPEED DISPLAYED

 20:1

15–2,000

15–2000

 1:1

 1,000–40,000 

1E3–40E3

• RATIO SETTING:

 

Press the number (1 or 20) that appears 

between

 the  

 and 

 buttons to select the ratio setting:

 RATIO SETTING

NUMBER DISPLAYED

 20:1

 20

 1:1

1

 

Press anywhere on the right side of the screen to return to the 

home screen.

 

TORQUE   —

 This segment of the screen displays the current 

torque setting. Press here   to open the torque adjustment 

screen, then press the 

 and 

 buttons to adjust the torque 

setting in the following ranges, according to the current ratio 

setting:

RATIO SETTING

TORQUE RANGE

 20:1

5–70 

 1:1

 1–4 

 

Press anywhere on the right side of the screen to return to the 

home screen.

 

The speed and ratio settings determine whether the system 

is in “

Torque-Limit

” mode or “

Torque Auto-Stop

” mode. 

See the table below for details.

TORQUE-LIMIT MODE 

When speed setting is 

≥500 RPM @ 20:1

 ratio

When speed setting is 

≥10,000 RPM @ 1:1

 ratio

A torque warning tone will beep when the handpiece 

torque reaches 75% of the torque setting. The handpiece 

will slow down when a load greater than the torque setting 

is reached. The handpiece will resume speed once the load 

is removed.

TORQUE AUTO-STOP MODE 

When speed setting is 

<500 RPM @ 20:1

 ratio

When speed setting is 

<10,000 RPM @ 1:1

 ratio

A torque warning tone will beep when the handpiece 

torque reaches 75% of the torque setting. The handpiece 

will stop one second after the handpiece reaches the torque 

setting. The handpiece will resume operation once the 

footpedal is released and reapplied.

 

Pressing the Torque screen segment 

 

also activates the 

Factory Restore and Calibration screens, as described in 

the sections below.

 

WATER   —

 This segment of the screen displays the current 

irrigation flow setting. Press here   to open the flow 

adjustment screen, then press the 

 and 

 buttons to adjust 

the flow setting (0–100%). A setting of 0% turns irrigation off. 

Press anywhere on the right side of the screen to return to the 

home screen.

Summary of Contents for AE-70V2

Page 1: ...U 1070 70V AEU 1070 Dental Implant Systems AE 70V2 Included with AEU 1070 70V AE 7PM Included with AEU 1070 AHP 85MB X Shown with optional equipment purchased separately OPERATION AND MAINTENANCE MANU...

Page 2: ...u mentation drills burs etc used in endodontics implantology and other oral surgery applications s CAUTION The irrigation supply system is designed for use with a saline solution or sterile water For...

Page 3: ...stant cabinet design helps prevent liquid spills from entering the internal electronics and allows easy clean up Isolated peristaltic pump chamber prevents leaks from entering the internal electronics...

Page 4: ...0 70V AEU 1070 Power Cord Console Cradle Motor and Cord AE 7PM Foot Controller AEU 1070 only AE 70V2 Foot Controller AEU 1070 70V only Spare Peristaltic Tubes Irrigation Tubing Set Bag Hanger PACKAGE...

Page 5: ...Connect the foot controller plug A for either the AE 7PM or AE 70V2 foot controller to the socket on the back of the console The foot controller plug is keyed to the socket and can only be inserted i...

Page 6: ...nd the small tubing connector E A C D E B 10 Two red arrows A on the pump door show the direction to install the irrigation tubing Start by threading the peristaltic tube segment through the top slot...

Page 7: ...03 clips B provided in the AE 24 tubing set bag A B 15 Hang a saline bag 1000ml max from the hanger with the saline port pointing towards the rear of the console Swing the bag away from the console to...

Page 8: ...ATIO SETTING Press the number 1 or 20 that appears between the and buttons to select the ratio setting RATIO SETTING NUMBER DISPLAYED 20 1 20 1 1 1 Press anywhere on the right side of the screen to re...

Page 9: ...d changes then stop 6 If calibration is successful the CAL button will stop flashing when the motor stops If the CAL button continues to flash after the motor stops calibration has failed Try calibrat...

Page 10: ...d Assembly CAUTION The motor is sensitive to shock Do not drop or impact the motor against a hard surface WARNING Do not oil or lubricate the motor CAUTION Do not allow oil from the handpiece to drain...

Page 11: ...utes B Pre vacuum Dynamic air removal method Time 4 minutes Temperature 132 C 270 F Dry time 40 minutes Flash Sterilization For immediate use only C Unwrapped standard autoclaving Gravity displacement...

Page 12: ...o setting matches the handpiece Check if handpiece lubricant is draining into motor Display error Switch off power wait 5 seconds then switch back on to reset Pump leaking Replace the peristaltic pump...

Page 13: ...n the fuse holder with the same type and rating 5 x 20mm 1 6A 250V slow blow 3 Reinstall the fuse holder SYMBOL DESCRIPTIONS Consult Instructions For Use Type B Equipment Caution Consult Accompanying...

Page 14: ...midity 10 to 90 non condensing Altitude 0 to 3 000 meters 0 to 9 842 feet Classifications Class 1 Equipment Type B Equipment Ordinary Equipment degree of protection against ingress of water Not suitab...

Page 15: ...PRST 1 PRST 2 PRST 3 PRST 4 PRST 5 PRST 6 PRST 7 PRST 8 PRESET DIRECTION SPEED RATIO TORQUE WATER PRST 1 PRST 2 PRST 3 PRST 4 PRST 5 PRST 6 PRST 7 PRST 8 PRESET DIRECTION SPEED RATIO TORQUE WATER PRS...

Page 16: ...548 Woodinville WA 98072 8333 216th Street S E Woodinville WA 98072 Phone 425 487 3157 Toll Free 800 426 5913 Web www aseptico com Email info aseptico com P N 421116 Rev C ECO 14662 02 2019 Printed in...

Reviews: