10

AEU-1070-70V / AEU-1070

AE-7PM Foot Controller

The AE-7PM foot controller provides basic on-off control of the 

system.

 

ON/OFF —

 Depress this footswitch to activate the motor and 

irrigation pump at the settings displayed on the screen.

MAINTENANCE AND STERILIZATION

Failure to adhere to the following instructions or blatant misuse of 

the device and/or related components will be considered abuse of the 

device and will void the warranty.

 

Call Aseptico at 1-800-426-5913 with any questions on 

sterilization procedures.

Handpiece

Follow the instructions provided with the handpiece for complete 

cleaning, maintenance and sterilization instructions.

Motor & Cord Assembly

!

 CAUTION

: The motor is sensitive to shock. Do not drop or 

impact the motor against a hard surface.

 WARNING

: Do not oil or lubricate the motor.

 CAUTION

: Do not allow oil from the handpiece to drain into 

the motor. After lubricating and before sterilizing the hand-

piece, stand it on a paper towel to allow excess oil to drain 

out of the handpiece.

 WARNING

: Do not attach a handpiece to the motor while 

the motor is running.

 WARNING

: Do not bend the motor cord sharply.

 WARNING

: Do not attempt to disassemble the motor or mo-

tor connector.

 WARNING

: Use of a sterilization method or temperatures 

other than what are prescribed below may damage com-

ponents or present a risk of cross-contamination between 

patients.

 CAUTION

: Do not soak or submerge the motor in any liquid.

 CAUTION

: Do not use ultrasonic cleaners.

 CAUTION

: Do not use cold sterilization or dry heat.

Summary of Contents for AE-70V2

Page 1: ...U 1070 70V AEU 1070 Dental Implant Systems AE 70V2 Included with AEU 1070 70V AE 7PM Included with AEU 1070 AHP 85MB X Shown with optional equipment purchased separately OPERATION AND MAINTENANCE MANU...

Page 2: ...u mentation drills burs etc used in endodontics implantology and other oral surgery applications s CAUTION The irrigation supply system is designed for use with a saline solution or sterile water For...

Page 3: ...stant cabinet design helps prevent liquid spills from entering the internal electronics and allows easy clean up Isolated peristaltic pump chamber prevents leaks from entering the internal electronics...

Page 4: ...0 70V AEU 1070 Power Cord Console Cradle Motor and Cord AE 7PM Foot Controller AEU 1070 only AE 70V2 Foot Controller AEU 1070 70V only Spare Peristaltic Tubes Irrigation Tubing Set Bag Hanger PACKAGE...

Page 5: ...Connect the foot controller plug A for either the AE 7PM or AE 70V2 foot controller to the socket on the back of the console The foot controller plug is keyed to the socket and can only be inserted i...

Page 6: ...nd the small tubing connector E A C D E B 10 Two red arrows A on the pump door show the direction to install the irrigation tubing Start by threading the peristaltic tube segment through the top slot...

Page 7: ...03 clips B provided in the AE 24 tubing set bag A B 15 Hang a saline bag 1000ml max from the hanger with the saline port pointing towards the rear of the console Swing the bag away from the console to...

Page 8: ...ATIO SETTING Press the number 1 or 20 that appears between the and buttons to select the ratio setting RATIO SETTING NUMBER DISPLAYED 20 1 20 1 1 1 Press anywhere on the right side of the screen to re...

Page 9: ...d changes then stop 6 If calibration is successful the CAL button will stop flashing when the motor stops If the CAL button continues to flash after the motor stops calibration has failed Try calibrat...

Page 10: ...d Assembly CAUTION The motor is sensitive to shock Do not drop or impact the motor against a hard surface WARNING Do not oil or lubricate the motor CAUTION Do not allow oil from the handpiece to drain...

Page 11: ...utes B Pre vacuum Dynamic air removal method Time 4 minutes Temperature 132 C 270 F Dry time 40 minutes Flash Sterilization For immediate use only C Unwrapped standard autoclaving Gravity displacement...

Page 12: ...o setting matches the handpiece Check if handpiece lubricant is draining into motor Display error Switch off power wait 5 seconds then switch back on to reset Pump leaking Replace the peristaltic pump...

Page 13: ...n the fuse holder with the same type and rating 5 x 20mm 1 6A 250V slow blow 3 Reinstall the fuse holder SYMBOL DESCRIPTIONS Consult Instructions For Use Type B Equipment Caution Consult Accompanying...

Page 14: ...midity 10 to 90 non condensing Altitude 0 to 3 000 meters 0 to 9 842 feet Classifications Class 1 Equipment Type B Equipment Ordinary Equipment degree of protection against ingress of water Not suitab...

Page 15: ...PRST 1 PRST 2 PRST 3 PRST 4 PRST 5 PRST 6 PRST 7 PRST 8 PRESET DIRECTION SPEED RATIO TORQUE WATER PRST 1 PRST 2 PRST 3 PRST 4 PRST 5 PRST 6 PRST 7 PRST 8 PRESET DIRECTION SPEED RATIO TORQUE WATER PRS...

Page 16: ...548 Woodinville WA 98072 8333 216th Street S E Woodinville WA 98072 Phone 425 487 3157 Toll Free 800 426 5913 Web www aseptico com Email info aseptico com P N 421116 Rev C ECO 14662 02 2019 Printed in...

Reviews: