background image

8

AEU-1000-70V / AEU-1000

CONTROL PANEL

This section describes the functions of the touch-screen.

Home Screen

The home screen displays the name of the active preset and the 

current settings for each of the preset’s adjustable parameters.

 

PRESET   —

 This segment of the screen displays the name of 

the current preset. Press here repeatedly to select presets 

sequentially. Below is a table of the factory settings for each 

preset:

PRESET

DIRECTION SPEED

RATIO

TORQUE

WATER

PRST-1

FORWARD

1200

20:1

50

70

PRST-2

FORWARD

800

20:1

50

70

PRST-3

FORWARD

15

20:1

35

0

PRST-4

FORWARD

15

20:1

50

0

PRST-5

REVERSE

25

20:1

50

0

PRST-6

FORWARD

20

20:1

30

0

PRST-7

FORWARD

21

20:1

40

0

PRST-8

FORWARD

22

20:1

50

0

 

MOTOR DIRECTION  —

 This segment of the screen displays 

the current motor rotation direction. Press here to alternate 

between forward (F) and reverse (R).

 

SPEED / RATIO   —

 This segment of the screen displays both 

the current speed 

and

 ratio settings. Press here to open the 

speed / ratio  adjustment  screen.

• SPEED SETTING: 

Press the 

 and 

 buttons to adjust the 

speed setting in the following ranges, according to the ratio 

setting: 

RATIO SETTING

SPEED RANGE

SPEED DISPLAYED

 20:1

15–2,000

15–2000

 1:1

 1,000–40,000 

1E3–40E3

 

RATIO INDICATOR

: If 

1:1  ratio

 is selected, the SPEED 

(RPM) will be displayed as a number ending in “

E3.

” 

Replace “E3” with “000” to calculate the speed of a 

1:1 handpiece (e.g. 40E3 = 40,000 RPM).  If 

20:1 ratio

 is 

selected, SPEED (RPM) is displayed as a whole number.

• RATIO SETTING: 

Press the number that appears 

between

  the  

 and 

 buttons to select the ratio setting:

NUMBER DISPLAYED

HANDPIECE RATIO SETTING

 20

 20:1

1

 1:1

 

Press anywhere on the right side of the screen to return to the 

home screen.

 

TORQUE   —

 This segment of the screen displays the current 

torque setting. Press here to open the torque adjustment 

screen, then press the 

 and 

 buttons to adjust the torque 

setting in the following ranges, according to the ratio setting:

RATIO SETTING

TORQUE RANGE

 20:1

5–50 

 1:1

 1–3 

 

Press anywhere on the right side of the screen to return to the 

home screen.

 

The  speed and ratio settings determine whether the 

system is in “

Torque-Limit

” mode or “

Torque Auto-

Stop

” mode. See the table below for details.

TORQUE-LIMIT MODE 

When speed setting is 

≥500 RPM 20:1

 ratio

When speed setting is 

≥10,000 RPM 1:1

 ratio

A torque warning tone will beep when the handpiece 

torque reaches 75% of the torque setting. The handpiece 

will slow down when a load greater than the torque setting 

is reached. The handpiece will resume speed once the load 

is removed.

TORQUE AUTO-STOP MODE 

When speed setting is 

<500 RPM 20:1

 ratio

When speed setting is 

<10,000 RPM 1:1

 ratio

A torque warning tone will beep when the handpiece 

torque reaches 75% of the torque setting. The handpiece 

will stop one second after the handpiece reaches the torque 

setting. The handpiece will resume operation once the 

footpedal is released and reapplied.

 

Pressing the Torque screen segment 

 

also activates the 

Factory Restore and Calibration screens, as described in 

the sections below.

 

WATER   —

 This segment of the screen displays the current 

irrigation flow setting. Press here to open the flow adjustment 

screen, then press the 

 and 

 buttons to adjust the flow 

setting (0–100%). A setting of 0% turns irrigation off. Press 

anywhere on the right side of the screen to return to the home 

screen.

Summary of Contents for AEU-1000

Page 1: ...AEU 1000 70V AEU 1000 Dental Implant Motor Systems AE 70V2 AE 7PM AHP 85MB X Shown with optional equipment purchased separately OPERATION AND MAINTENANCE MANUAL...

Page 2: ...u mentation drills burs etc used in endodontics implantology and other oral surgery applications s CAUTION The irrigation supply system is designed for use with a saline solution or sterile water For...

Page 3: ...0 70V or basic on off foot controller AEU 1000 Water resistant cabinet design helps prevent liquid spills from entering the internal electronics and allows easy clean up Isolated peristaltic pump cham...

Page 4: ...0 70V AEU 1000 Power Cord Console Cradle Motor and Cord AE 7PM Foot Controller AEU 1000 only AE 70V2 Foot Controller AEU 1000 70V only Spare Peristaltic Tubes Irrigation Tubing Set Bag Hanger PACKAGE...

Page 5: ...4 Connect the foot controller plug A for either the AE 7PM or AE 70V2 foot controller to the socket on the back of the console The foot controller plug is keyed to the socket and can only be inserted...

Page 6: ...nd the small tubing connector E A C D E B 10 Two red arrows A on the pump door show the direction to install the irrigation tubing Start by threading the peristaltic tube segment through the top slot...

Page 7: ...the console immediately A 16 Press the power switch A on the rear of the console to the on I position to turn the unit on A Handpiece Selection Screen 17 The handpiece selection screen is displayed e...

Page 8: ...R DISPLAYED HANDPIECE RATIO SETTING 20 20 1 1 1 1 Press anywhere on the right side of the screen to return to the home screen TORQUE This segment of the screen displays the current torque setting Pres...

Page 9: ...ough several speed changes then stop 6 If calibration is successful the CAL button will stop flashing when the motor stops If the CAL button continues to flash after the motor stops calibration has fa...

Page 10: ...d Assembly CAUTION The motor is sensitive to shock Do not drop or impact the motor against a hard surface WARNING Do not oil or lubricate the motor CAUTION Do not allow oil from the handpiece to drain...

Page 11: ...utes B Pre vacuum Dynamic air removal method Time 4 minutes Temperature 132 C 270 F Dry time 40 minutes Flash Sterilization For immediate use only C Unwrapped standard autoclaving Gravity displacement...

Page 12: ...o setting matches the handpiece Check if handpiece lubricant is draining into motor Display error Switch off power wait 5 seconds then switch back on to reset Pump leaking Replace the peristaltic pump...

Page 13: ...n the fuse holder with the same type and rating 5 x 20mm 1 6A 250V slow blow 3 Reinstall the fuse holder SYMBOL DESCRIPTIONS Consult Instructions For Use Type B Equipment Caution Consult Accompanying...

Page 14: ...midity 10 to 90 non condensing Altitude 0 to 3 000 meters 0 to 9 842 feet Classifications Class 1 Equipment Type B Equipment Ordinary Equipment degree of protection against ingress of water Not suitab...

Page 15: ...PRST 1 PRST 2 PRST 3 PRST 4 PRST 5 PRST 6 PRST 7 PRST 8 PRESET DIRECTION SPEED RATIO TORQUE WATER PRST 1 PRST 2 PRST 3 PRST 4 PRST 5 PRST 6 PRST 7 PRST 8 PRESET DIRECTION SPEED RATIO TORQUE WATER PRS...

Page 16: ...548 Woodinville WA 98072 8333 216th Street S E Woodinville WA 98072 Phone 425 487 3157 Toll Free 800 426 5913 Web www aseptico com Email info aseptico com P N 421100 Rev E ECO 15016 06 2020 Printed in...