10
AEU-1000-70V / AEU-1000
AE-7PM Foot Controller
The AE-7PM foot controller provides basic on-off control of the
system.
ON/OFF —
Depress this footswitch to activate the motor and
irrigation pump at the settings displayed on the screen.
MAINTENANCE AND STERILIZATION
Failure to adhere to the following instructions or blatant misuse of
the device and/or related components will be considered abuse of the
device and will void the warranty.
Call Aseptico at 1-800-426-5913 with any questions on
sterilization procedures.
Handpiece
Follow the instructions provided with the handpiece for complete
cleaning, maintenance and sterilization instructions.
Motor & Cord Assembly
!
CAUTION
: The motor is sensitive to shock. Do not drop or
impact the motor against a hard surface.
WARNING
: Do not oil or lubricate the motor.
CAUTION
: Do not allow oil from the handpiece to drain into
the motor. After lubricating and before sterilizing the hand-
piece, stand it on a paper towel to allow excess oil to drain
out of the handpiece.
WARNING
: Do not attach a handpiece to the motor while
the motor is running.
WARNING
: Do not bend the motor cord sharply.
WARNING
: Do not attempt to disassemble the motor or mo-
tor connector.
WARNING
: Use of a sterilization method or temperatures
other than what are prescribed below may damage com-
ponents or present a risk of cross-contamination between
patients.
CAUTION
: Do not soak or submerge the motor in any liquid.
CAUTION
: Do not use ultrasonic cleaners.
CAUTION
: Do not use cold sterilization or dry heat.