FUSE REPLACEMENT

 

 

WARNING

 

Switch off the power and unplug the unit  

before following the steps below.

 

The system features auto-sensing global voltage compatibility. 

The fuse indicated is correct for 100V-240V 50/60 Hz line 

voltage. 

1 . 

Remove the fuse holder from the power inlet socket by prying 

it out using two small blade screwdrivers, one on the top 

A

 

and the other on the bottom 

B

 of the fuse holder.

A

B

2 . 

Replace both fuses in the fuse holder with the same type and 

rating: 

5 x 20mm, 1 .6A, 250V slow-blow

.

3 . 

Reinstall the fuse holder.

SYMBOL DESCRIPTIONS

 

 

Consult Instructions For Use

 

 

Type B Equipment

 

 

Caution – Consult Accompanying Documents

 

 

Alternating Current

 

 

Fuse Rating

 

 

Dangerous Voltage

 

 

Motor Direction

 

 

Preset Step Through

 

 

Torque Step Through

 

 

Pump On-Off/Step Through

 

 

Dispose Of Properly

 

 Footswitch

 

 Manufacturer

 

 

Ingress of Water - Classification-Footswitch

 

 

Protective Earth (Ground)

 

 

Temperature Limitation

 

 

Atmospheric Pressure Limitation

 

 

Humidity Limitation

 

 

Sterilize at 132°C (270°F)

 

  Authorized European Representative

 

 

Serial Number

 

 

Part Number

 

 

ON-OFF Switch

 

 

Important information

Summary of Contents for AEU-1000

Page 1: ...AEU 1000 70V AEU 1000 Dental Implant Motor Systems AE 70V2 AE 7PM AHP 85MB X Shown with optional equipment purchased separately OPERATION AND MAINTENANCE MANUAL...

Page 2: ...u mentation drills burs etc used in endodontics implantology and other oral surgery applications s CAUTION The irrigation supply system is designed for use with a saline solution or sterile water For...

Page 3: ...0 70V or basic on off foot controller AEU 1000 Water resistant cabinet design helps prevent liquid spills from entering the internal electronics and allows easy clean up Isolated peristaltic pump cham...

Page 4: ...0 70V AEU 1000 Power Cord Console Cradle Motor and Cord AE 7PM Foot Controller AEU 1000 only AE 70V2 Foot Controller AEU 1000 70V only Spare Peristaltic Tubes Irrigation Tubing Set Bag Hanger PACKAGE...

Page 5: ...4 Connect the foot controller plug A for either the AE 7PM or AE 70V2 foot controller to the socket on the back of the console The foot controller plug is keyed to the socket and can only be inserted...

Page 6: ...nd the small tubing connector E A C D E B 10 Two red arrows A on the pump door show the direction to install the irrigation tubing Start by threading the peristaltic tube segment through the top slot...

Page 7: ...the console immediately A 16 Press the power switch A on the rear of the console to the on I position to turn the unit on A Handpiece Selection Screen 17 The handpiece selection screen is displayed e...

Page 8: ...R DISPLAYED HANDPIECE RATIO SETTING 20 20 1 1 1 1 Press anywhere on the right side of the screen to return to the home screen TORQUE This segment of the screen displays the current torque setting Pres...

Page 9: ...ough several speed changes then stop 6 If calibration is successful the CAL button will stop flashing when the motor stops If the CAL button continues to flash after the motor stops calibration has fa...

Page 10: ...d Assembly CAUTION The motor is sensitive to shock Do not drop or impact the motor against a hard surface WARNING Do not oil or lubricate the motor CAUTION Do not allow oil from the handpiece to drain...

Page 11: ...utes B Pre vacuum Dynamic air removal method Time 4 minutes Temperature 132 C 270 F Dry time 40 minutes Flash Sterilization For immediate use only C Unwrapped standard autoclaving Gravity displacement...

Page 12: ...o setting matches the handpiece Check if handpiece lubricant is draining into motor Display error Switch off power wait 5 seconds then switch back on to reset Pump leaking Replace the peristaltic pump...

Page 13: ...n the fuse holder with the same type and rating 5 x 20mm 1 6A 250V slow blow 3 Reinstall the fuse holder SYMBOL DESCRIPTIONS Consult Instructions For Use Type B Equipment Caution Consult Accompanying...

Page 14: ...midity 10 to 90 non condensing Altitude 0 to 3 000 meters 0 to 9 842 feet Classifications Class 1 Equipment Type B Equipment Ordinary Equipment degree of protection against ingress of water Not suitab...

Page 15: ...PRST 1 PRST 2 PRST 3 PRST 4 PRST 5 PRST 6 PRST 7 PRST 8 PRESET DIRECTION SPEED RATIO TORQUE WATER PRST 1 PRST 2 PRST 3 PRST 4 PRST 5 PRST 6 PRST 7 PRST 8 PRESET DIRECTION SPEED RATIO TORQUE WATER PRS...

Page 16: ...548 Woodinville WA 98072 8333 216th Street S E Woodinville WA 98072 Phone 425 487 3157 Toll Free 800 426 5913 Web www aseptico com Email info aseptico com P N 421100 Rev E ECO 15016 06 2020 Printed in...

Reviews: