background image

2

SETTING UP THE UNIT:

1.  Plug the AEU-10 console into a

standard 120 Volt AC grounded power

outlet. (220 Volt AC unit is also

available)

2.  Plug the micro motor into the

receptacle on the front of the console

labeled 'motor'. Make sure the power

switch on the console is turned off.

3.  Attach the straight handpiece to the

motor by sliding straight onto the

motor shaft. Insert a bur or bit into the

handpiece collet. Only operate the

unit with a bit or blank installed and

with the collet in the closed position.

4.  Attach AE-7P on/off foot pedal to the

receptacle on the back of the unit

marked “Foot Switch”. The AE-7P foot

pedal is standard with the AEU-10SS

and optional with the AEU-10C. The

variable speed AE-14A foot control is

also available.

Motor

Handpiece

Bit

Console

Motor Holder

Figure 1

Foot

Pedal

On / Off

Switch

‘ON’

Indicator

Foot Switch

Receptacle

AC Power

Inlet

Motor

AHP-101

Handpiece

Bit

(Close Collet)

Fuse Holder

Summary of Contents for AEU-10C/SS

Page 1: ...AEU 10C SS Power Tip Systems OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Page 2: ...nored NOTE Important additional information and hints WARNING This device has been tested and found to comply with the emissions requirements of EN 55014 1997 These requirements provide reasonable pro...

Page 3: ...ase read the instructions provided in this manual to receive the best and longest service from your Aseptico equipment Separate manuals may be provided to cover the operation and maintenance of handpi...

Page 4: ...aft Insert a bur or bit into the handpiece collet Only operate the unit with a bit or blank installed and with the collet in the closed position 4 Attach AE 7P on off foot pedal to the receptacle on t...

Page 5: ...nd selector must be moved to Hand to return to hand mode Figure 2 POWER SWITCH ON INDICATOR HAND FOOT SELECTOR SPEED DIAL MOTOR RECEPTACLE FWD REV SELECTOR NOTE If motor stops due to an open collet or...

Page 6: ...turning grip 1 4 turn as shown in Fig 4 until you feel a click 2 Push bit back into collet and turn to release 3 Pull bit straight out of collet Bit Collet Grip Gap Retainer Close Collet Insert Bur Op...

Page 7: ...foot pedal connection if used Check motor plug connection Test for possible bad handpiece by removing handpiece from motor and then running motor by itself Check that Hand Foot Selector switch is plac...

Page 8: ...ty is at its sole option and discretion to repair or replace any defective component or product in part or whole Aseptico shall be the sole arbiter of such action In the event of alleged defect under...