4

To Install Bit:

1. 

Open collet by turning grip 1/4 turn as

shown in 

Fig. 4

until you feel a

“click”. The gap between the grip and

handpiece retainer will narrow when

the collet is open. 

2.

Insert bit fully into collet.

3.

Close collet by turning grip 1/4 turn

as shown in 

Fig. 3

NOTE:

Before use, check to insure that bit is

fully inserted and grip is firmly secured

in locked position.

HANDPIECE:

The Aseptico AEU-10 System includes the AHP-101 straight handpiece.  This handpiece

features a quick-twist chuck that will accept any standard HP (3/32") shank bit (or bur)

or any Doriot style attachment.

To Remove Bit:

1.

Open collet by turning grip 1/4 turn

as shown in 

Fig. 4

until you feel a

“click”.

2.

Push bit back into collet and turn to

release.

3.

Pull bit straight out of collet.

Bit   Collet               Grip            Gap     Retainer

Close Collet

Insert Bur

Open Collet

Push, Turn,

and Pull Bur

Sterilization:

The AHP-101 steam autoclaving sterilization procedure provided below has been

independently validated to achieve a sterility assurance level (SAL) of 10

-6

.

This handpiece must be sterilized before initial use and between each patient use per

the following procedure:

Scrub dirt and other debris from handpiece 

and wipe clean with a cotton swab or

cloth.

•  

Autoclave

at 132°C (270°F) for ten minutes.

•  

Do not rinse or submerge

in any solution, or use ultrasonic cleaners.

•  

Do not lubricate.

The sealed bearings in this handpiece eliminate the need for

lubrication.

Figure 3

Figure 4

Summary of Contents for AEU-10C/SS

Page 1: ...AEU 10C SS Power Tip Systems OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Page 2: ...nored NOTE Important additional information and hints WARNING This device has been tested and found to comply with the emissions requirements of EN 55014 1997 These requirements provide reasonable pro...

Page 3: ...ase read the instructions provided in this manual to receive the best and longest service from your Aseptico equipment Separate manuals may be provided to cover the operation and maintenance of handpi...

Page 4: ...aft Insert a bur or bit into the handpiece collet Only operate the unit with a bit or blank installed and with the collet in the closed position 4 Attach AE 7P on off foot pedal to the receptacle on t...

Page 5: ...nd selector must be moved to Hand to return to hand mode Figure 2 POWER SWITCH ON INDICATOR HAND FOOT SELECTOR SPEED DIAL MOTOR RECEPTACLE FWD REV SELECTOR NOTE If motor stops due to an open collet or...

Page 6: ...turning grip 1 4 turn as shown in Fig 4 until you feel a click 2 Push bit back into collet and turn to release 3 Pull bit straight out of collet Bit Collet Grip Gap Retainer Close Collet Insert Bur Op...

Page 7: ...foot pedal connection if used Check motor plug connection Test for possible bad handpiece by removing handpiece from motor and then running motor by itself Check that Hand Foot Selector switch is plac...

Page 8: ...ty is at its sole option and discretion to repair or replace any defective component or product in part or whole Aseptico shall be the sole arbiter of such action In the event of alleged defect under...

Reviews: