Page 18
Size: . . . . . . . . . . . . . . . . . . . . . . . . . .17.5” x 11.6” x 6.9”
(44.5 cm x 29.5 cm x
17.5 cm)
Volume: . . . . . . . . . . . . . . . . . . . . . . .0.81 ft
3
(0.23 m
3
)
Weight: . . . . . . . . . . . . . . . . . . . . . . .27 lbs
(12.25 kg)
Power Source: . . . . . . . . . . . . . . . . . .AC Dual Voltage
120V/230V @ 50Hz/60Hz
12V/24V vehicle battery
27.6V battery pack (x 2)
Water Reservoir Capacity: . . . . . . . .16 fl. oz. (0.423 liter)
Air Reserve Capacity, Standard: . . .25.8 in
3
(423 cc)
Noise Levels . . . . . . . . . . . . . . . . . . .64 dBa @ 1 meter (3’4”)
Fuses: . . . . . . . . . . . . . . . . . . . . . . . .120V: 1.25A, 250V slo-blo fuse
(Installed at factory)
230V: 0.6A, 250V slo-blo fuse
Fuse Size: . . . . . . . . . . . . . . . . . . . . .5 x 20mm
Duty Cycle for AE-200-30 Motor: . .Intermittent use:
1 minute On /
3 minutes Off
Mains Current Ratings: . . . . . . . . . .120V: 1.25A, 230V: 0.6A
NOTE: The appliance inlet is the mains disconnect means.
Environmental Conditions:
- Operating Temperature: 2° to 49° C (35° to 120° F)
- Transport/Storage Temperature: -20° to 60° C (-4° to 140° F)
- Relative Humidity: 5 to 95% non-condensing
- Altitude: 0 to 2591 meters (0 to 8,500 feet)
NOTE: Air pressure instruments are set to 2591 meters (8,500 ft) elevation.
WARNING:
Equipment not suitable for use in the presence of a
flammable anaesthetic mixture with air, or with oxygen or nitrous
oxide.
CAUTION:
If the integrity of the protective earth ground is in doubt,
only use battery pack or vehicle battery power for operation.
• Type of Protection Against Electric Shock:
-
Class I when system is powered by mains power source.
-
Class II when system is powered by Vehicle Battery
power source.
-
Internally Powered Equipment for Battery Pack
power source.
• Degree of Protection Against Electric Shock: Type B Applied Part
• Degree of Protection Against Ingress of Water: Ordinary Equipment
• Mode of Operation: Continuous
SPECIFICATIONS
MEDICAL EQUIPMENT WITH RESPECT
TO ELECTRIC SHOCK, FIRE AND
MECHANICAL HAZARDS ONLY IN
ACCORDANCE WITH UL 2601-1
CAN/CSA C22.2 No. 601.1
This device has been tested and found to comply with the emissions requirements of IEC
60601-1-2:2001-09. These requirements provide reasonable protection against harmful
electromagnetic interference in a typical medical installation. However, high levels of radio-
frequency (RF) emissions from electrical devices, such as cellular phones, may disrupt the
performance of this device. To mitigate disruptive electromagnetic interference, position this
device away from RF transmitters and other sources of electromagnetic energy.
IPX1
IPX1
Alternating current
Protective earth (ground)
Protect Against Dripping Water
Dangerous Voltage
Attention, consult
accompanying documents
SYMBOL DEFINITIONS:
Type B Equipment
Do Not Throw Into Trash