4
QUICK START OPERATION
After the unit has been set up, you are ready
for quick start operation as follows:
1. Turn the Power Switch on the rear panel of
the console from the 'OFF' (0) to the
'ON' ( –) position.
The Console LCD Display will light and the
Startup Screen (shown below) will be
displayed for a few seconds. The Startup
Screen displays the software version
onboard your Endo DTC
®
.
ASEPTICO AEU-25
Software Rev. ASP05
2. Insert an endodontic rotary file into the
handpiece.
3. Use the Calibration function to calibrate the
Endo DTC
®
. See Caution at right.
Press and release the “CAL” button. The
unit will run for a few seconds, measuring
the handpiece for an accurate display of
speed and torque values.
4. Select the Preset Button corresponding to
the endodontic rotary file series you are
using.
By default, the Endo DTC
®
preset buttons
are set for the file series shown in Chart 1.
Pressing the Preset Button repeatedly will
cycle through the individual files listed
below each series name.
5. Press the foot pedal to activate motor and
begin operation.
When using the optional AE-7D Dual Foot
Control:
The black (right side) pedal controls the
on/off operation of the handpiece (exactly
like the standard AE-7P foot pedal).
The blue (left side) pedal provides a hands-
free way to operate the preset buttons:
• To step to the next individual file of an
entire file series on the active preset
button: Press and release the blue pedal.
• To step to the next preset button: Press
and hold the blue pedal for 1/2 sec. (until
you hear a double beep), then release
the blue pedal.
6. After 20 minutes of inactivity, the Endo
DTC
®
will go into Sleep Mode to conserve
power. To exit Sleep Mode and resume
operation, press the Foot Pedal.