9.
10. Motor Handpiece Light
- To operate the
handpiece light, turn on the light switch,
adjust the light intensity by turning in a
clockwise position, and press the foot
pedal. After the foot pedal is released, the
light will turn off after approximately 10
seconds. The light switch also acts a
dimmer. Rotating the light knob changes
the intensity of the light. If the user flips the
Scaler/Handpiece Toggle Switch to the
scaler position, the fiber optic light will
automatically turn off.
11. Foot Switch:
The Foot Switch provides on/off operation
to the Motor and water coolant to the
handpiece when the Scaler/Motor Toggle
Switch is on the right (Motor) side. The
Foot Switch provides on/off operation to
the Optional Ultrasonic Scaler and water
coolant to the Scaler Instrument when the
Scaler/Motor Toggle Switch is on the left
(Scaler) side.
12. Water Supply Bottle
:
The Aseptico AEU-425FO incorporates a
self contained pressurized water system.
This consists of a 0.75-liter bottle
dispensing water through the 3-Way
Air/Water Syringe, for Handpiece Coolant,
and for Ultrasonic Scaler Coolant. The
Water Supply Bottle attaches to the
threaded reservoir connector, located
below the pressure gauge in the AEU-
425FO case.
Do not store water in the Water Supply
Bottle while transporting the AEU-425FO.
To refill the Water Supply Bottle:
1. Before removing the bottle for refilling, it
must be depressurized. Turn the Water
Supply On/Off Toggle into the "Off"
position, allowing pressure to release.
2. Unscrew bottle
3. Fill with water.
4. Screw bottle onto reservoir connector
5. Turn the Water Supply On/Off Toggle to
the “On” position.
13. Air Reserve Tank
:
The air reserve tank is located behind the
water supply tank and stores pressurized
air for the 3-Way Air/Water Syringe. The
Air Reserve tank is easily removed to
drain condensation.
14. HVE and Saliva Ejector Vacuums
:
The Transport II is equipped with a High
Volume (HVE) and a Low Volume (Saliva
Ejector) Vacuum System. Both vacuum
systems operate from separate vacuum
sources with automatic controls.
When the Vacuum Holder Toggle Switch is
off (to the left), the vacuum system will
stay off even when the vacuum valves are
removed from the holder until the toggle
WARNING
:
Before unscrewing the Water Supply
Bottle, vent reserve pressure. Turn the
Water Supply On/Off Toggle into the
"Off" position, allowing pressure to
release.
WARNING
:
Before unscrewing the Air Reserve
Tank, vent reserve air pressure in the
tank. Press the right (Air) button on the
3-Way Air/Water Syringe or open the
low vacuum, with unit power off, to
depressurize the air reserve tank.