9.

10.  Motor Handpiece Light 

- To operate the

handpiece light, turn on the light switch,

adjust the light intensity by turning in a

clockwise position, and press the foot

pedal. After the foot pedal is released, the

light will turn off after approximately 10

seconds. The light switch also acts a

dimmer. Rotating the light knob changes

the intensity of the light. If the user flips the

Scaler/Handpiece Toggle Switch to the

scaler position, the fiber optic light will

automatically turn off.

11. Foot Switch:

The Foot Switch provides on/off operation

to the Motor and water coolant to the

handpiece when the Scaler/Motor Toggle

Switch is on the right (Motor) side. The

Foot Switch provides on/off operation to

the Optional Ultrasonic Scaler and  water

coolant to the Scaler Instrument when the

Scaler/Motor Toggle Switch is on the left

(Scaler) side.

12. Water Supply Bottle

:

The Aseptico AEU-425FO incorporates a

self contained pressurized water system.

This consists of a 0.75-liter bottle

dispensing water through the 3-Way

Air/Water Syringe, for Handpiece Coolant,

and for Ultrasonic Scaler Coolant. The

Water Supply Bottle attaches to the

threaded reservoir connector, located

below the pressure gauge in the AEU-

425FO case.
Do not store water in the Water Supply

Bottle while transporting the AEU-425FO.

To refill the Water Supply Bottle:

1. Before removing the bottle for refilling, it

must be depressurized. Turn the Water

Supply On/Off Toggle into the "Off"

position, allowing pressure to release.

2. Unscrew bottle 
3. Fill with water.
4. Screw bottle onto reservoir connector
5. Turn the Water Supply On/Off Toggle to

the “On” position.

13. Air Reserve Tank

:

The air reserve tank is located behind the

water supply tank and stores pressurized

air for the 3-Way Air/Water Syringe. The

Air Reserve tank is easily removed to

drain condensation.

14. HVE and Saliva Ejector Vacuums

:

The Transport II is equipped with a High

Volume (HVE)  and a Low Volume (Saliva

Ejector) Vacuum System. Both vacuum

systems operate from separate vacuum

sources with automatic controls.
When the Vacuum Holder Toggle Switch is

off (to the left), the vacuum system will

stay off even when the vacuum valves are

removed from the holder until the toggle

WARNING

Before unscrewing the Water Supply

Bottle, vent reserve pressure. Turn the

Water Supply On/Off Toggle into the

"Off" position, allowing pressure to

release.

WARNING

Before unscrewing the Air Reserve

Tank, vent reserve air pressure in the

tank. Press the right (Air) button on the

3-Way Air/Water Syringe or open the

low vacuum, with unit power off, to

depressurize the air reserve tank.

Summary of Contents for AEU-425 Transport II

Page 1: ...AEU 425FO Transport II Portable Dental System With Fiber Optics OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Page 2: ...eptico com info aseptico com To prevent injury to people and damage to property please heed relevant warnings and remarks They are marked as follows WARNING Serious injury or death may result if ignor...

Page 3: ...many years of reliable service Please read the instructions provided in this manual to receive the best and longest service from your Aseptico equipment Separate manuals may be provided to cover the o...

Page 4: ...Reserve Capacity 45 8 cu in 750 cc Water Reservoir Capacity 25 4 fl oz 0 75 liter Water Flow 6 5 fl oz min 0 192 liters min High Volume Waste Tank Capacity 0 2 gal 0 75 L Low Volume Waste Tank Capaci...

Page 5: ...NG Do not install where there is a risk of an explosion The Transport II is not intended for operation in the presence of flammable anesthetics or gases WARNING CONTRAINDICATIONS The AEU 425FO Transpo...

Page 6: ...motor See Figs 1 2 3 4 Attach the supplied AE 7P On Off Foot Switch to the connector on the front of the Electric Control Console marked Foot Switch 5 Move the 3 Way Air Water Syringe from the travel...

Page 7: ...MENT HOLDER DETAIL FULL WASTE CONTAINER INDICATOR LIGHT 5 10 Place the small Waste Tank in the unit with the small gray connector towards the front Attach the gray hose coming from above to the white...

Page 8: ...PTACLE ELECTRIC CONTROL CONSOLE SCALER HANDPIECE TOGGLE SWITCH OPTIONAL SCALER COOLANT FLOW ADJUSTMENT OPTIONAL SCALER ON OFF SWITCH HANDPIECE COOLANT FLOW ADJUSTMENT HANDPIECE COOLANT ON OFF SWITCH O...

Page 9: ...INER 3 WAY SYRINGE STORAGE COMPRESSOR INTAKE AIR FILTER BUNGEE CORDS COMPRESSOR COOLING FAN WATER SUPPLY ON OFF TOGGLE VACUUM HOLDER TOGGLE SWITCHES PRESSURE GAUGE FULL WASTE CONTAINER INDICATOR LIGHT...

Page 10: ...andpiece Speed Ratio Range RPM 1 5 Increaser 25 000 150 000 1 1 4 000 30 000 4 1 Reduction 1 000 7 500 16 1 Reduction 300 1 800 clockwise to decrease water flow and counterclockwise to increase water...

Page 11: ...in the AEU 425FO case Do not store water in the Water Supply Bottle while transporting the AEU 425FO To refill the Water Supply Bottle 1 Before removing the bottle for refilling it must be depressuriz...

Page 12: ...g large debris upon vacuum intake 14 Waste Level Indicator When either waste container is full the red LED will light Vacuum and pressure operation will cease until the container is emptied To empty t...

Page 13: ...ssing the foot switch and allowing the water to flow for 30 seconds 3b Screw on a sterile Instrument Tip to the Scaler Handpiece and tighten it firmly Turn carefully backwards approx 1mm This will put...

Page 14: ...nd of motor housing and insert dental pick or other small pointed instrument into guide hole at base of bulb see photograph Slide bulb forward until disconnected from socket in motor housing Wipe bulb...

Page 15: ...hose with a soft cloth Use a 45 isopropal and detergent solution DO NOT IMMERSE the handpiece in any fluid or spray any fluid directly on the handpiece WATER LINES Disinfect the water lines weekly Pre...

Page 16: ...as far as it will go Release collar and gently tug on tip before using to ensure that tip is securely locked into socket Syringe Tip Sterilization 1 Remove contaminated syringe tip 2 Remove all visibl...

Page 17: ...t connector 2 Changing the fuse Replacement Fuses 0 63A 250V slo blo fuses Fuse size 5 x 20mm 3 Replace the Fuseholders EXTENSION CORD RECOMMENDATIONS Use only a 3 wire extension cord that has a 3 bla...

Page 18: ...tention consult accompanying documents SYMBOL DEFINITIONS Type B Equipment Protect Against Dripping Water Serial Number IPX1 Scaler Adjustment Switches Toggle Select Switch Foot Switch Handpiece Coola...

Page 19: ...into motor After lubricating set handpiece with head down to let excess lubricant drain out Check bulb Note Do Not Handle Bulb With Fingers Ensure that HVE and Low Volume Saliva Ejector vacuum heads...

Page 20: ...option and discretion to repair or replace any defective component or product in part or whole Aseptico shall be the sole arbiter of such action In the event of alleged defect under warranty the purch...

Reviews: