When the Vacuum Holder Toggle Switch
is in the on position (to the right), the
vacuum will start automatically when the
vacuum valves are removed from the
holder. The vacuum pump will start and
waste from the vacuum system will collect
in the plastic waste containers. The waste
container should be emptied when it is ¾
full or when the red LED full waste
container light is lit.
The High Volume (HVE) Vacuum System
contains a screen inside the Waste Tank
Container. This is used for collecting large
debris upon vacuum intake.
14. Waste Level Indicator
:
When either waste container is full, the
red LED will light. Vacuum and pressure
operation will cease until the container is
emptied.
To empty the Waste Container:
1. Turn off the unit.
2. Disconnect hoses.
3. Disconnect electrical level connection.
(Pull on connector, not wire)
4. Remove front bungee cord.
5. Empty container.
6. Reconnect hoses and connector.
OPERATION FUNCTIONS (continued):
10.
IMPORTANT NOTE:
To prevent the HVE and Low Volume
(Saliva Ejector) compressor system
from running excessively, ensure that
the vacuum heads are firmly cradled in
their holders after each use and that
their spring-loaded control levers are
toggled ‘Off’’.