13.
VACUUM SySTEM
:
The HVE and Saliva Ejector Valves are fully
autoclavable. Remove the valves from the
hoses before autoclaving.
The vacuum hoses should not be autoclaved.
Clean with a disinfectant solution.
CAUTION
:
Use only NON-foaming cleansers in the
vaccum lines.
OPTIONAL ULTRASONIC SCALER
:
See accompanying documents.
The Scaler Handpiece Cover and Instrument
Tips are fully autoclavable.
Disinfect and clean the Scaler Handpiece
Cover and Instrument Tips before autoclaving.
Autoclave at a maximum temperature of 135º
C (275º F) for 10 minutes or 120º C (248º F)
for 20 minutes.
Wipe off the Scaler Handpiece and it’s
silicone hose with a soft cloth. Use a 45%
isopropal and detergent solution.
DO NOT
IMMERSE
the handpiece in any fluid or spray
any fluid directly on the handpiece.
WATER LINES
:
Disinfect the water lines weekly. Prepare a
1:10 bleach solution (1 part household bleach
to 10 parts water). Remove water reservoir
and discard residual water. Replace empty
water supply tank and air purge all waterlines.
Fill water supply tank with bleach solution.
Run bleach solution through all lines. Allow
bleach solution to stand in lines for 10
minutes. Remove water supply tank and
discard bleach. Flush water supply tank and
all lines thoroughly with clean water. Air purge
and leave lines dry until next clinical use..
CONSOLE
:
The exterior of the console may be cleaned by
wiping with a soft cloth moistened with mild
detergent or disinfecting solution.