20
Problem:
Console does not light when
turned on:
Console lights when turned on,
but handpiece does not turn:
No water flow from pump to
handpiece:
Irrigation tube leaks:
Correction:
• Check console to power connection.
• Check fuse. If blown, replace with 1.6A/250V
slo-blow fuse.
• Check motor plug connection.
• Check foot switch connection.
• Depress foot switch.
• Increase RPM.
• Increase Torque setting
• Check that bur/drill is properly seated in the
handpiece and the collet is closed.
• Check that pump is on and flow level is sufficient.
• Check that water container seal is completely
punctured.
• Make sure the irrigation tubing is properly installed
in pump door and flow is in the correct direction.
• Replace worn tube section located under the pump
door with a new section from extra tube set
provided with this system.
TROUBLESHOOTING:
CHANGING THE FUSE:
1.
Remove the Fuse Holder from the Power
Inlet connector (see Figure 17).
2.
Replace the fuses in the Fuse holder.
Replacement Fuses:
1.6A, 250V slo-blow fuse
(Fuse size: 5 x 20mm)
3.
Reinstall the Fuse Holder.
Turn the power off and unplug the unit
before following the steps below.
WARNING
Fig. 17 - FUSES
NOTE:
The AEU-7000 & AEU-7000-70V
feature auto-sensing, global voltage
compatibility. The fuse indicated is correct for
100V-240V 50/60 Hz line voltage.