MAINTENANCE & CLEANING
:
SPECIFICATIONS:
:
Console Dimensions:
9.98"W x 9.42"L x 5.10"H
(25.3 cm x 23.9 cm x 12.9 cm)
Console Weight:
7.3 lbs (3.3 kg)
Power:
100-240V
1.1 - 0.5 A
50-60 Hz
Fuses:
1.6A, 250V, Slo-Blow Type
Duty Cycle:
16.7%
NOTE:
The appliance inlet is the mains disconnect means.
Environmental Conditions:
• Operating Temperature: 10° to 28°C (50° to 82.4°F)
• Transportation & Storage Temperature:
• -20° to 60°C (-4° to 140°F)
• Relative Humidity: 10% to 90% noncondensing
• Altitude: 0 to 3,048 meters (0 to 10,000 feet)
MEDICAL ELECTRIC EQUIPMENT WITH
RESPECT TO ELECTRIC SHOCK, FIRE
AND MECHANICAL HAZARDS ONLY IN
ACCORDANCE WITH UL 60601-1 (First
Edition) and CAN/CSA C22.2 No. 601.1-M90
41EJ
WARNING: This device has been tested and found to comply
with the emissions requirements of IEC 60601-1-2:2001-09.
These requirements provide reasonable protection against
harmful electromagnetic interference in a typical medical
installation. However, high levels of radio-frequency (RF)
emissions from electrical devices, such as cellular phones, may
disrupt the performance of this device. To mitigate disruptive
electromagnetic interference, position this device away from RF
transmitters and other sources of electromagnetic energy.
10°C
(50°F)
28°C
(82.4°F)
OPERATING
TEMPERATURE
-20°C
(-4°F)
60°C
(140°F)
TRANSPORT &
STORAGE
TEMPERATURE
1013.3
hPa
697
hPa
10%
90%
HANDPIECES
- Thorough cleaning and lubrication of
handpieces after each use and before sterilization is very
important to ensure proper operation and service life of the
handpiece. Follow the instructions provided with the
handpiece for complete maintenance instructions.
MOTOR
-
IMPORTANT!
Protect motor
from excess oil draining from handpiece.
After lubricating and before autoclaving,
stand handpiece by its base on a paper
towel and allow excess oil to drain (see
Figure 40).
AEU-7000L MOTOR LED LENS CLEANING
- The lens of
the LED light on the AEU-7000L
motor (see Fig. 41) is soft and
can be damaged. It should not
be exposed to dust and debris.
Excessive dust and debris may
cause a drastic decrease in
optical output. In the event that
the light requires cleaning, first
try a gentle swabbing, using a
lint-free swab. If needed, use a lint-free swab and isopropyl
alcohol to gently remove dirt from the lens.
Do not use
other solvents
as they may adversely react with the LED
assembly.
CONSOLE
- The exterior of the console may be cleaned
by wiping with a soft cloth moistened with a mild
detergent or a 1:10 bleach solution (1 part household
bleach to 10 parts water).
IMPORTANT:
Use of other
cleaning or disinfecting solutions may damage the
console and may void the warranty.
SILICONE WATER LINES
- The silicone water lines used
for the pump are fully autoclavable:
Pre-Cleaning:
Before sterilization, run clean water
through the tubing for 30 seconds to expel any stagnant
water.
NOTE:
Do not use disinfectants on the tubing set.
Bacteria and viruses will be neutralized during sterilization.
Sterilization:
Sterilize tubing at 132º C (270º F) for 10
minutes.
FOOT CONTROL
- The exterior of the foot control may
be cleaned by wiping with a soft cloth moistened with
mild detergent or disinfecting solution. When cleaning,
remove handle from foot control and wipe clean with
disinfectant, then reinstall handle.
• Do not attempt to disassemble the motor or motor connector.
• Do not oil or lubricate the motor.
• Do not attach a handpiece to the motor while the motor is
running.
• Do not bend motor cord sharply.
• The motor is sensitive to shock. Do not drop or impact motor
against a hard surface.
WARNING
Failure to comply with any of the above instructions may void your warranty.