20
The
SETUP
Program allows the
user to select/configure setup
options via display prompts. The
option selections and corresponding
instructions are shown below:
1.
Press and hold the SETUP/CANCEL
button to enter SETUP Mode. The
following prompt will be displayed:
Recall Factory Setup?
Press: YES / NO / CANCEL
a.
Press “Yes” to recall the factory setup
menu. The following prompt will display:
Are You Sure?
Press: YES / NO / CANCEL
b.
To return the System preset buttons to
their factory default settings, press YES.
Important:
Any customized presets will
be lost when factory settings are recalled.
c.
To continue with System Setup (and
keep all customized settings), press NO.
d.
A brief message announces which Setup
Mode,
Implant
or
Endo
, is being
activated.
Setup Only For
Implant (or Endo) Mode
2.
The user is prompted next to enable the
Torque Warning Tone feature. This feature
warns the user with an audible signal
when a specified Torque Limit is reached.
Warning signals are provided differently,
depending upon which operating mode is
activated. When in
Implant
Mode, a
separate signal is emitted when the torque
reaches each of the two major thresholds:
1.) 75% of specified Torque Limit - Emits a
fast beeping signal.
2.) 100% of specified Torque Limit - Emits
a slow beeping signal.
When in
Endo
Mode, a single signal is
emitted when the torque reaches the
specified limit:
1.) 100% of specified Torque Limit - Emits
a fast beeping signal.
The following prompt will display:
Torque Warning Tone?
Press: YES / NO / CANCEL
a.
To enable the Warning Tone, press Yes.
b.
To disable the Warning Tone, press No.
3.
The next prompt provides a Reverse
Warning Tone that alerts the user
whenever the handpiece is rotating in the
reverse (counterclockwise) direction:
Reverse Warning Tone?
Press: YES / NO / CANCEL
a.
To enable the Reverse Tone, press Yes.
b.
To disable the Reverse Tone, press No.
NOTE:
The Reverse Warning Tone emits a
beeping signal with a slow 1/2-second
cadence that is easily distinquishable from
the two Torque Warning Tones in Step #2
above.
IMPORTANT:
If both Torque and Reverse
Warning Tone options are enabled and
activated at the same time (e.g., user
reaches 75% of torque while running in
reverse), the Torque warning tones will
override the Reverse warning — only the
Torque signal will be heard.
SETUP OPTIONS - Implant & Endo
Description
1. Recall Factory Setup
2. Torque Warning Tone
3. Reverse Warning Tone
4. Auto Stop Mode
Endodontic Mode Only
5. Sleep Mode
6. Variable Speed Pedal
7. Save Your Settings
SYSTEM SETUP
Summary of Contents for AEU-7000E
Page 31: ...29 NOTES...