4. Auto Stop Mode (for Endo Mode Only):
When the Auto Stop Reverse (ASR)
feature is enabled, rotation of the
handpiece will automatically stop and
reverse when the selected torque limit is
reached. Do not release the foot pedal
when this occurs. Keep the foot pedal
activated and the Endo System will
automatically alternate between forward
and reverse rotation in an attempt to free
the instrument.
When the Auto Stop Manual (ASM) feature
is enabled, forward rotation of the
handpiece will automatically stop when the
selected torque limit is reached. Upon
releasing and then reapplying pressure to
the foot control, the handpiece will rotate
in the reverse direction. The handpiece will
continue to rotate in reverse until the foot
control is released once again. If pressure
is then reapplied to the foot control, the
handpiece will return to forward rotation.
The following prompt will display:
Auto Stop Mode:
1=ASR 2=ASM / CANCEL
a.
To enable Auto-Stop Reverse (ASR),
press Preset Button #1 .
b.
To enable Auto-Stop Manual (ASM), and
disable ASR, press Preset Button #2.
5.
The next prompt provides a choice of two
different time delays before the System
enters Sleep Mode, wherein the Display
and Keypad time out and become inactive:
Sleep Mode
1=15 2=30 3=Off / CANCEL
a.
To enable a 15-minute delay, press
Preset Button #1.
b.
To enable a 30-minute delay, press
Preset Button #2.
c.
To disable Sleep Mode (Display stays
on), press Preset Button #3.
NOTE: The system default is the 30-minute
delay.
6.
This prompt allows the user to choose
whether the Variable Speed Foot Control
operates in Variable Mode (‘0’ to ‘set’
speed), or in On/Off Mode (runs only at
‘set’ speed):
Variable Pedal Mode
1=Var 2=On/Off / CANCEL
a.
To enable Variable (“Var”) Mode, press
Preset Button #1.
b.
To enable “On/Off” Mode, press Preset
Button #2.
NOTE: The system defaults are Variable
when in
Implant
Mode, and On/Off when
in
Endo
Mode.
7.
This final prompt asks if the new settings
are to be saved:
Save Your Settings ?
Press: YES / NO
a.
To save new settings, press YES.
b.
To discard your new settings and keep
the previous settings, press NO.
NOTE:
The System will automatically Exit
the “SETUP” menu at the completion of
this Step.