25
STERILIZATION:
WARNING
- Sterilize the motor between
each patient use.
WARNING
- Use of a sterilization
method or temperatures other than what
are prescribed may damage the motor
or present a risk of cross-contamination
between patients.
CAUTION
- Do not soak or submerge
the motor in any liquid.
STERILIZATION PROCEDURE:
Pre-clean
1)
Brush off any visible signs of debris from
the motor and cord.
2)
Thoroughly clean the device with a moist
cloth or towel to remove any remaining signs
of debris.
Sterilize
3)
Select one of the three following
sterilization methods (A. B. or C.):
Wrapped Sterilization
– Place in an
appropriately sized sterilization pouch and
seal it.
A. Standard autoclaving (Gravity displacement
method)
Time:
15 min
Temperature:
132° C (270° F)
Dry time:
30 minutes
B. Pre-vacuum (dynamic-air-removal)
Time:
4 minutes
Temperature:
132° C (270° F)
Dry time:
40 minutes
Flash Sterilization
– For immediate use
only.
C
.
Unwrapped standard autoclaving (Gravity
displacement method)
Time:
10 minutes
Temperature:
132° C (270° F)
No dry time is required for flash sterilization.
Motor & Cord Assembly:
The entire AE-230M-40 motor and cord
assembly is fully autoclavable. Loosely coil
the motor cord when autoclaving. Avoid
sharply bending the cord when autoclaving.
NOTE:
Call Aseptico Inc. at 1-800-426-5913
for any questions or clarifications on this
sterilization procedure.
The entire motor &
cord assembly is
autoclavable.
MOTOR & CORD STERILIZATION
Fig.15