3
WARNING:
Always comply with the handpiece and implant/file manufacturers’ instructions
regarding maximum speeds, torques, forward and reverse directions, and use of all
instrumentation, drills, burs, etc., used in endodontics, implantology, and other oral
surgery applications.
CAUTION:
The irrigation supply system is designed for use with a saline solution or sterile
water. For implants, use only suitable irrigants as recommended by the
manufacturer’s instructions.
CAUTION:
Connect mains power cable to a properly grounded outlet only.
CAUTION:
The motor is sensitive to shock and may be damaged if dropped or impacted
against a hard surface.
WARNING:
Do not disassemble or alter the System motor, console, or foot switch.
CAUTION:
Use only appliance cord Type C13,10A per IEC / EN 60320-1. Note: North America,
Denmark, Australia, and New Zealand may require hospital grade plugs. Consult
local codes.
WARNING:
Never use damaged or worn files as they may separate in the root canal.
WARNING:
Do not use this device in conjunction with an electric scalpel or on patients with
pacemakers.
CAUTION:
Never connect or disconnect the bag spike to the irrigation bag over the console.
Water spilled onto the console can damage the unit.
CAUTION:
It is recommended to always have the patient wear a rubber dam during endodontic
procedures.
CAUTION:
Not every implant contra angle is rated up to 80 Ncm of torque. Before using this
motor, contact your handpiece supplier to verify the appropriate torque range of the
implant contra angle(s) you intend to use on this device. Do not adjust the torque
above the supplier-recommended rating or there is a risk of damage to the internal
parts of your handpiece (which is not covered under warranty). Aseptico
recommends and distributes AHP-85MB-series 20:1 handpieces, which are rated
up to 80Ncm.