4
SPEED RANGES
1/1
4,000 - 30,000 RPM
1/16
300 - 1,900
RPM
1/20
200 - 1,500
RPM
1/ 256 10 - 120
RPM
Fig. 5b - AEU-707AV2 Speed Ranges
OPERATION FUNCTIONS:
1. Power Switch
- Controls power On / Off to
console. When turned on, console should
light up.
2. Pump On / Off
- Activates irrigation pump.
When turned on, the flow display will light
and the pump will activate when either
motor A or B is turned on.
3. Pump Flow
- Increases or decreases the
irrigation flow. Push the Up/Down arrow to
increase flow or decrease flow. A visual dis-
play shows the approximate volume level.
4. RPM
- Controls speed of motor and hand-
piece. Push the Up arrow to increase speed
and the Down arrow to decrease speed. A
visual display shows actual RPM. Ratio
selector must match exactly the ratio of the
hand piece being used for complete accura-
cy.
5. Forward / Reverse
- Determines the rotation
of the motor and handpiece. A red LED light
displays the reverse rotation.
6. Ratio Selector
- Calibrates the RPM display
to read accurately for up to four different
handpiece ratios. Push the ratio Up / Down
arrow until the indicator lights match the
handpiece ratio you are using. Different
handpiece ratios are required to obtain high
torque, low speeds used in surgery. The
speed ranges with a 30K motor are shown
in Figures 5a (AEU-707A) and 5b (AEU-
707AV2).
Fig. 5a - AEU-707A Speed Ranges
SPEED RANGES
1/1
4,000 - 30,000 RPM
1/16
300 - 1,900
RPM
1/64
50 - 470
RPM
1/ 256 10 - 120
RPM