STERILIZATION:

WARNING

Sterilize the motor

between each patient use.

WARNING 

Use of a sterilization

method or temperatures other than what
are prescribed may damage the motor
or present a risk of cross-contamination
between patients.

CAUTION 

Do not soak or submerge

the motor in any liquid.

STERILIZATION PROCEDURE:
Pre-clean

1)

Thoroughly brush off any visible signs of

debris from the motor and cord.

2)

Thoroughly clean the device with a moist

cloth or towel to remove any remaining signs
of debris.

Sterilize

3)

Select one of the three following

sterilization methods (A. B. or C.):

Wrapped Sterilization

– 

Place in an

appropriately sized sterilization pouch and
seal it.

A. Gravity Wrapped:

• Minimum Temperature:

132° C (270° F) 

• Full Cycle Time:

20 minutes 

• Minimum Dry time:

30 minutes 

B. Prevacuum Wrapped: 

• Preconditioning Pulses: 

• Minimum Temperature:

132° C (270° F)

• Full Cycle Time: 

4 minutes 

• Minimum Dry time:

40 minutes 

Flash Sterilization

– For immediate use

only.

C

Gravity Unwrapped:

• Minimum Temperature:

132° C (270° F) 

• Full Cycle Time:

10 minutes 

No dry time is required for flash
sterilization.

MOTOR & CORD ASSEMBLY:

The entire motor and cord assembly is fully
autoclavable. Before autoclaving the
assembly, remove any handpiece from motor

and install autoclaving plugs into the end of
the motor and motor cord connector as
shown below. The plugs are supplied with the
motor and cord.

Loosely coil the motor cord when auto-
claving.  Avoid sharply bending the cord when
autoclaving.

NOTE:

Call Aseptico, Inc. at 1-800-426-5913

for any questions or clarifications on this
sterilization procedure.

•  Always install the autoclaving plugs into motor

and motor connector before autoclaving.

•  If the autoclaving plug for the motor becomes

difficult to remove or install, apply silicone grease
or similar lubricant to the O-rings.

Turn black connector ring only

HANDPIECES

Thorough cleaning and

lubrication of handpieces after each use and
before sterilization is very important to ensure
proper operation and service life of the
handpiece. Follow the instructions provided with
the handpiece purchased for complete
maintenance instructions.

CONSOLE MAINTENANCE

The exterior of

the console may be cleaned by wiping with a
soft cloth moistened with mild detergent or
disinfecting solution.

IRRIGATION TUBING SETS - 

Irrigation tubing

sets can be removed from the pump and may
be sterilized by autoclave, chemclave, or
ethylene oxide gas processes. Follow the
manufacturer’s recommended procedure.

NEVER store the tubing sets in the pump for
extended periods of time. The tubing may
collapse and stick to itself creating a blockage
for future uses.

MAINTENANCE & CLEANING:

• Do not attempt to disassemble the motor or motor connector.
• Do not oil or lubricate the motor.
• Do not attach a handpiece to the motor while motor is running.
• Do not bend motor cord sharply.

WARNING

Failure to comply with any of the above instructions may

void your warranty

6

Summary of Contents for AEU-707A

Page 1: ...AEU 707A AEU 707AV2 Implant Surgery Systems OPERATION AND MAINTENANCE INSTRUCTION MANUAL ...

Page 2: ...LASSIFICATIONS Internally Powered Equipment Type B Equipment Ordinary Equipment degree of protection against ingress of water Not suitable for use in the presence of a flam mable anesthetic mixture with air or with oxy gen or nitrous oxide P O Box 1548 Woodinville WA 98072 8333 216th Street S E Woodinville WA 98072 425 487 3157 800 426 5913 www aseptico com info aseptico com WARNING This device ha...

Page 3: ...iable service Please read the instructions provided in this manual to receive the best and longest service from your Aseptico equipment Separate manuals may be provided to cover the operation and maintenance of hand held pieces or other accessories for your unit PACKAGE CONTENTS Electronic Control Console AE 4B 30 Autoclavable Brushless Micro Motor Autoclavable Motor Holder AE 7P On Off Foot Contr...

Page 4: ... is shipped with the console 2 Attach the remote power cord to the back of the console and plug into a grounded elec trical receptacle 3 Connect the motor to the receptacle on the lower front panel of the console 4 Attach the appropriate E Type handpiece to the motor as shown in Figure 3 5 Insert bag hanger rod into holder on the top of the unit Thread into place 6 Attach the supplied foot control...

Page 5: ...ound the rollers with slight tension and insert into bottom slot on the pump block d While maintaining tension on the tubing check that tubing is centered on the rollers and is completely tucked into slots on pump block Then replace clear pump cover by sliding into position from front to back and securing with the thumbscrew e Route the remaining length of tubing to the handpiece and connect to th...

Page 6: ...Up arrow to increase speed and the Down arrow to decrease speed A visual display shows actual RPM Ratio selector must match exactly the ratio of the hand piece being used for complete accura cy 5 Forward Reverse Determines the rotation of the motor and handpiece A red LED light displays the reverse rotation 6 Ratio Selector Calibrates the RPM display to read accurately for up to four different han...

Page 7: ...h the pump to the ON position The display will light Set the flow to the desired level Thorough cleaning and lubrication of hand pieces after each use and before steriliza tion is very important to ensure proper operation and service life of the handpiece Follow the instructions provided with the handpiece purchased for complete mainte nance instructions NOTE The RPM display indicates true bur spe...

Page 8: ...otor and cord Loosely coil the motor cord when auto claving Avoid sharply bending the cord when autoclaving NOTE Call Aseptico Inc at 1 800 426 5913 for any questions or clarifications on this sterilization procedure Always install the autoclaving plugs into motor and motor connector before autoclaving If the autoclaving plug for the motor becomes difficult to remove or install apply silicone grea...

Page 9: ... If blown replace with 1A 250V slo blo fuse for 110 volts 0 5A 250V slo blo fuse for 220 volts Check motor plug connection Check foot switch connection Depress foot switch Increase RPM Check that bur is properly seated in the handpiece and the collet is closed Check that pump is on and flow level is sufficient Check that water container seal is completely punctured Make sure the irrigation tubing ...

Page 10: ...8 NOTES ...

Page 11: ...NOTES ...

Page 12: ...tion In the event of alleged defect under warranty the purchaser is to notify Aseptico s Customer Service Department promptly Customer Service will provide instructions usually directing that the product be returned for service Shipment to Aseptico and the cost thereof is always the responsibility of the purchaser Accidental misuse inappropriate installation or failure to perform directed maintena...

Reviews: