background image

4

3.

Attach power cord to receptacle on
bottom of power supply housing (see
Figure 2, page 3) and plug cord into a
grounded electrical outlet. 

NOTE:

The

ALU-40LED Lamp features auto-sensing,
global voltage compatibility for 100VAC,
115VAC, 230VAC, and 240VAC at 50Hz
and 60Hz. The fuses provided are correct
for these line voltages. 

NOTE:

A USA

style 110V power cord and 220V adapter
are shipped with the unit. Power plug
adapters for the specific country of usage
may need to be purchased separately.

Optional DC Power:

If the power source is 12VDC or 24VDC,
use the optional DC Battery Clip or
Adapter Assemblies (see Figure 3) to
connect to the DC source. Insert the
Assemblies' Power Plugs into the DC
Power Jack located on the bottom of the
power supply housing (see Figure 2, page
3).

NOTE:

When attaching the optional DC

Battery Clips to the battery, connect the
red cable clamp to the positive (+)
terminal, and the black clamp to the
negative (-) terminal. 

SETTING UP THE UNIT

- Cont'd

Fig. 3 - Optional DC Power Accessories 

DC POWER PLUG

ASSEMBLY

(Optional)

DC BATTERY CLIP

ASSEMBLY

(Optional)

Summary of Contents for ALU-40LED

Page 1: ...OPERATION AND MAINTENANCE INSTRUCTION MANUAL ALU 40LED PORTABLE DENTAL LAMP ALU 40LED PORTABLE DENTAL LAMP...

Page 2: ...al 425 487 3157 Toll Free 800 426 5913 www aseptico com info aseptico com INFORMATION CONCERNING THE ACCURACY AND PRECISION OF THIS PRODUCT MAY BE OBTAINED UPON REQUEST BY CONTACTING ASEPTICO AT THE A...

Page 3: ...ded in this manual to receive the best and longest service from your Aseptico equipment Separate manuals may be provided to cover the operation and maintenance of other accessories for your unit PACKA...

Page 4: ...ed for operation in the presence of flammable anesthetics or gases CAUTION Connect AC power supply mains cable to a properly grounded outlet only CAUTION The ALU 40LED Lamp is sensitive to shock and m...

Page 5: ...ired height Tighten the handgrip to secure post in place Loosen the boom adjustment handle and extend boom to desired length and or pivot the lamp head to the desired operating position Tighten handle...

Page 6: ...ific country of usage may need to be purchased separately Optional DC Power If the power source is 12VDC or 24VDC use the optional DC Battery Clip or Adapter Assemblies see Figure 3 to connect to the...

Page 7: ...ble brightness control When viewed from the rear of the Lamp Head rotate Ring clockwise to increase lamp intensity or counter clockwise to decrease intensity The intensity adjustment range is from 0 t...

Page 8: ...r shelf on the Front Housing see Figure 7 e Attach Lens to Housing with the three Fasteners from Step 1c above f Replace Lens Cover 2 LENS ADJUSTMENT Follow these steps to adjust the position of the L...

Page 9: ...sh soap to a lint free lens tissue and gently wipe Lens Clean the Lens Cover using the same method Reinstall Fresnel Lens and Lens Cover onto Lamp Head b Metal Components The metal exterior may be cle...

Page 10: ...8 ALU 40LED PARTS DIAGRAM Figure 9 Lamp Stand Boom ALU 40LED LAMP HEAD Refer to Figure 11 page 12 1 2 3 4 5 6 7 9 11 10 8 12 13 5 14 11...

Page 11: ...SOC M6 X 8MM BLK 510731 1 HEAT SHRINK 1 8 BLK 870118 1 HEAT SHRINK 1 4 BLK 870120 2 5 HOLE PLUG NYLON 1 2 IN BLACK 510202 1 M S STNLS PHDPHL 10 24 X 5 8 THRD CUTTING 510732 1 END CAP MODIFIED BOOM AL...

Page 12: ...10 ALU 40LED PARTS DIAGRAM Figure 10 Lamp Stand Power Supply 15 16 20 22 23 24 41 19 17 37 21 16 33 36 28 35 34 26 27 28 23 25 32 31 29 30 40 38 39 42 18...

Page 13: ...0419 1 TERM INSUL RING 6 22 16 860004 1 WIRE PVC 22 GRN YEL 54 870260 0 21 CONN JACK DC POWER PANEL MOUNT 860258 1 FUSE 5 X 20 SLO BLO 0 160 AMP 830046 2 FUSE DRAWER 2 POLE 840060 1 POWER INLET FUSED...

Page 14: ...12 Figure 11 Lamp Head 2 4 5 8 9 18 17 6 3 6 7 10 11 13 19 16 15 1 21 12 20 14...

Page 15: ...1 WASHER SHIM 43 OD X 115 ID X 01 STNLS 510726 1 APERTURE ADJUST RING ALU 40LED 461931 1 RES 100K TRIM POT 350V 1 2W 810430 1 WIRE 24 TEFLON 41 40 TINNED COPPER WHT 870314 2 WIRE 24 TEFLON 41 40 TINNE...

Page 16: ...up Replace fuses 0 160 Amp 5 x 20mm 250V slo blow Check plug and electrical source Turn intensity up Clean Lens Cover and or Fresnel Lens Note Before cleaning turn power to Lamp Off Check plug and el...

Page 17: ...er 5 72 cm to 11 74 cm Adjustable 27 56 in 70 cm focal length Beam Characteristics Color temperature 6 500 Kelvin Light Output 1 500 ft c 24 Typ 16 145 lux 61 cm Typ Light Distribution 7 different ape...

Page 18: ...tion consult accompanying documents Type B Equipment Do Not Throw Into Trash Class II Equipment Manufacturer Authorized European Representative Protective Earth Ground Temperature Limitation Atmospher...

Page 19: ...ter of such action In the event of alleged defect under warranty the purchaser is to notify Aseptico s Customer Service Department promptly Customer Service will provide instructions usually directing...

Page 20: ...P N 420841 Rev F ECO 13305 03 2014 PRINTED IN THE USA P O Box 1548 Woodinville WA 98072 8333 216th Street S E Woodinville WA 98072 425 487 3157 800 426 5913 www aseptico com info aseptico com...

Reviews: