AMC‑15
MAINTENANCE AND STERILIZATION
1.
WASTE REMOVAL —
Empty
the waste tank daily.
Waste removal steps:
a.
Open the waste compart-
ment door on the rear of
the unit.
b.
Remove the end of waste
hose from tethered hose
clamp
A
, then direct the
waste hose to a suitable
drain.
c.
Switch ON the waste purge pump
B
to empty the waste
tank. It may take a minute before waste water begins to
flow from the end of the hose.
d.
Return the end of the waste hose into the clamp
A
, then
pinch clamp closed to prevent leakage.
The type of optional amalgam separator used with the
AMC-15 requires no settlement time before emptying the
waste tank.
2.
WASTE SYSTEM CLEANING —
After emptying the waste
tank each day, the waste system should be cleaned.
The waste system consists of:
• Waste tank
• Waste discharge hose
• HVE (high volume ejector) valve and vacuum line
• Saliva ejector (low volume) valve and vacuum line
Cleaning steps:
a.
Prepare 1 liter of a 1:10 bleach solution (1 part household
bleach to 9 parts water).
b.
Remove the disposable ends from both the HVE and saliva
ejector valves.
c.
Vacuum half of the bleach solution through the HVE valve,
then the other half through the saliva ejector valve by
putting the tip of each valve in the bleach solution.
d.
Flush each valve and line by vacuuming 1/2 liter of clean
water through each valve.
e.
Empty the waste tank following the instructions in step 1
above.
f.
Clean the outside of the ejector valves and hoses with a
disinfectant solution.
3.
AMALGAM SEPARATOR
—
The
optional
amalgam
separator
A
is located in the
waste compartment. The
separator filters out particu-
late matter while the waste
water is being pumped out of
the unit. A significant drop in
waste discharge pressure will
occur when the amalgam sep-
arator becomes full of solids.
Replacement amalgam sep-
arators (PN 730595) are avail-
able through Aseptico. Refer
to the label on the amalgam
separator for the address of
an approved amalgam recy-
cler. Replace the separator
by unscrewing the amalgam
separator retaining knob
B
located under the top lid,
and removing the amalgam
separator from the waste
compartment. NOTE: You can
operate the AMC-15 with the
bypass filter
C
installed in
place of the amalgam separa-
tor, where permitted.
4.
AMALGAM SOLIDS TRAP —
Replace the disposable
amalgam solids trap routinely. The performance of the
vacuum system will decrease significantly as the trap fills
up with solids. The trap sits inside the central vacuum canis-
ter located under the top lid. Keep spare traps in stock. The
AMC-15 uses standard 1 7/8” traps (both open and closed type
traps are compatible):
OPEN TYPE CLOSED TYPE
PART NO. QUANTITY
PART NO. QUANTITY
AA-29
1
AA-290VS
1
AA-29-100
100
AA-290VS-100
100
NOTE:
Check with state laws regarding the recycling of
amalgam solids waste.
5.
AIR TANK PURGE —
Empty
the air tank daily to remove
condensation from air tank.
Place a container underneath
the air tank drain hose
A
to
catch water. Rotate the air
tank pressure knob
B
count-
er-clockwise to empty the air tank, then rotate knob clockwise
to close valve when done.
A
B
B
A
C
B
A