13
AMC‑15
TROUBLESHOOTING GUIDE
Unit will not start:
•
Check system power connection.
•
Check circuit breakers.
•
Check if waste tank sensor is connected.
•
Check if waste tank is full.
Unit starts but trips circuit breakers:
•
Check source circuit to see if it is a minimum of 15A.
•
NOTE: The use of extension cords is not recommended.
No water pressure:
•
Check water bottles for water level and verify that both caps
are tight.
•
Check that water bottle pressure toggle is in the ON position.
Insufficient vacuum:
•
Check vacuum hose assemblies for blockage.
•
Check amalgam solids trap under the top lid for blockage.
•
Check that the waste tank lid is properly seated and tightly
secured.
Insufficient handpiece operation:
•
Check the pressure gauge on the front of the cabinet and
ensure that sufficient air is being delivered to the handpiece.
•
Check that handpiece tubing is untangled and not crimped.
•
Check handpiece connection for missing gasket.
•
Check the system pressure gauge under the top lid. System
pressure is factory set and should be approximately 80 psi (5.5
bar). Call aseptico customer service if system pressure needs
to be adjusted.
No water to handpiece:
•
Check that the toggle on the foot control is in the ON position.
•
Check that bottle select toggle is selecting a bottle filled with
water (or that both bottles are full).
•
Check that the bottle pressure toggle is in the ON position.
•
Check that the water coolant valve to the selected handpiece
is open.
No coolant air to handpiece:
•
Check that the air coolant toggle is in the ON position.
•
Check that the air coolant flow control valve is open.
Waste pump is not working:
•
Check the circuit breaker next to the waste pump switch.
•
Check to see if the amalgam separator is clogged by replacing
it with the bypass filter.
•
Test if the check valve to the amalgam separator is clogged by
removing the tube from the check valve closest to the separa-
tor and seeing if a cup of water can be extracted.
Unit is turning On and Off:
•
Check if waste tank sensor is firmly connected.
•
Check if waste tank is full.
Amalgam filter or bypass filter leaks:
•
Check that the O-rings are properly installed onto the ends of
fittings.
MESSAGES
SYSTEM MESSAGES
None
ERROR MESSAGES
None
FAULT MESSAGES
None
REPLACEMENT PARTS LIST
The following items can be purchased from Aseptico for replace-
ment (or as spares) of standard and optional equipment.
PART NO.
ITEM
QTY
730595
FILTER REPLACEMENT AMALGAM SEPARATOR
1
730631-01
BOTTLE 1000ML ZIRC KIT MODIFIED
(CLEAR 1-LITER WATER BOTTLES W/ LEVEL INDICATORS)
2
AA-20A
AUXILIARY ARM WITH ACCESSORY TRAY
1
AA-29
SOLIDS COLLECTOR DISPOSABLE SCREEN 1 7/8” SINGLE
(AMALGAM SOLIDS TRAP (OPEN TYPE))
1
AA-29-100
SOLIDS COLLECTOR DISPOSABLE SCREEN 1 7/8” SINGLE
PKG 100
(AMALGAM SOLIDS TRAP (OPEN TYPE))
100
AA-290VS
SOLIDS COLLECTOR DISPOSABLE COVERED 1 7/8” SINGLE
(AMALGAM SOLIDS TRAP (CLOSED TYPE))
1
AA-290VS-100
SOLIDS COLLECTOR DISPOSABLE COVERED 1 7/8” SINGLE
PKG 100
(AMALGAM SOLIDS TRAP (CLOSED TYPE))
100
TA-1
SYRINGE TIP AUTOCLAVABLE 3-WAY
1