14
AMC‑15
SYMBOL DEFINITIONS
Caution – Consult Accompanying Documents
Consult Instructions For Use
Do Not Use If Damaged
Use By Date
Part Number
Follow instructions for use
Type B Applied Part
Warning—Potential danger to patient or user
(consult accompanying documents)
Dangerous Voltage
Alternating Current
This symbol indicates that the waste of electri-
cal and electronic equipment must not be dis-
posed as unsorted municipal waste and must
be collected separately.
Air Coolant Control
Water Coolant Control
Air Coolant Toggle -or- Flush Toggle
Authorized European Representative
Motor Direction
Protective earth (ground)
Protect Against Dripping Water
Serial Number
Footswitch
On/Off Switch - Auxiliary
On/Off Switch - Mains
Bottle Pressure Release Toggle
Bottle Select Toggle
Manufacturer
Atmospheric Pressure Limitation
Temperature Limitation
Humidity Limitation
Sterilizable up to the temperature indicated
Waste Tank full -or- Waste Level Sensor Cable
disconnected
SPECIFICATIONS
Cart Size
23 5” W x 30 0” L x 36 5” H (56 7 cm x 76 2 cm x 92 7 cm)
Cart Weight (with all optional equipment)
139 lb (63 kg)
Shipping Crate Size
30 0” W x 37 0” L x 50 0” H (76 2 cm x 93 9 cm x 127 cm)
Shipping Crate Weight
72 lb (32 7 kg)
Power Source
115V version:
115V @ 60Hz
230V version:
230V @ 50Hz
Power Rating
115V version:
11 1A @ 60Hz, 115VAC
230V version:
6 5A @50Hz, 230VAC
Circuit Breakers
115V version:
Input: 15A, Pressure: 8A, Vacuum: 8A, Waste Pump: 1A & 3/4A
230V version:
Input: 15A, Pressure: 5A, Vacuum: 5A , Waste Pump: 1A & 3/4A
Operating Pressure
80 psi (5 5 bar)
High Volume Vacuum
6 0 SCFM @ 0 in Hg (169 9 liters/min @ 0 cm Hg)
3 6 SCFM @ 4 in Hg (101 9 liters/min @ 10 2 cm Hg)
Low Volume Vacuum
1 3 SCFM @ 1 5 in Hg (36 8 liters/min @ 4 in Hg)
Pressure Pump
1 8 SCFM @ 100 psi (51 0 liters/min @ 6 9 bar)
Air Storage Capacity
2 08 gal (7 9 liters) nominal
Air Storage Pressure
100 psi (6 9 bar)
Water Bottle Capacity
67 8 fl oz (2 0 liters)
Water Flow
5 07 fl oz /min (0 15 liter/min)
Waste Tank Capacity
1 08 gal (4 08 liters)
Noise Level
Compressor: 58 dBA @ 3’4” (1 meter)
Vacuum: 62 dBA @ 3’4” (1 meter)
Compressor & Vacuum Combined: 63 dBA @ 3’4” (1 meter)
AMC-15 Cart Duty Cycle
Continuous
Environmental Conditions
Operating Temperature: 50° to 104° F (10° to 40° C)
Transport/Storage Temperature: -40° to 160° F (-40° to 71° C)
Relative Humidity: 10 to 95% non-condensing
Altitude: 0 to 10,000 feet (0 to 3048 meters)
Summary of Contents for AMC-15
Page 1: ...OPERATION MAINTENANCE MANUAL AMC 15 Mobile Dental Cart...
Page 17: ...17 AMC 15 NOTES...
Page 18: ...18 AMC 15...