18.
OPTIONAL MOTOR LED LENS CLEANING:
The lens of the LED light on the optional electric
motor (Fig. 33) is soft and can be damaged. It
should not be exposed to dust and debris.
Excessive dust and debris may cause a drastic
decrease in optical output. In the event that the
light requires cleaning, first try a gentle
swabbing, using a lint-free swab. If needed,
use a lint-free swab and isopropyl alcohol to
gently remove dirt from the lens.
Do not use
other solvents
as they may adversely react
with the LED assembly.
OPTIONAL ULTRASONIC SCALER
:
(
Refer to ultrasonic scaler instructions for
use, supplied separately.)
The scaler handpiece cover and scaler tips are
fully autoclavable. Disinfect and clean the cover
and tips before autoclaving. Autoclave at a
maximum temperature of 135º C (275º F) for
10 minutes or 120º C (248º F) for 20 minutes.
Wipe off the scaler handpiece and its silicone
hose with a soft cloth. Use a 45% isopropyl and
detergent solution.
DO NOT IMMERSE
the
handpiece in any fluid or spray any fluid directly
on the handpiece.
OPTIONAL CURING LIGHT
:
(
Refer to curing light instructions for use,
Sunlite Lazer model, supplied separately.
Sunlite Lazer is a product of Kinetic Instruments, Inc.
)
The curing probe is fully autoclavable. Detach
probe by pressing quick disconnect button on
side of handle. Clean probe with disinfectant
then autoclave at 135°C (275°F) for 20 minutes
minimum. Clean and sterilize probe between
each patient use.
NOTE:
The lamp module
(the section that seats into the tubing
connector) is NOT autoclavable.
OPTIONAL ELECTRIC MOTOR/CORD
RECEPTACLE O-RINGS
The O-rings for the three water/air ports in the
optional electric motor/cord receptacle should
be replaced if damaged or worn. Use the
provided O-ring installer pin and sleeve to
replace the O-rings:
1. Remove old O-ring from water or air port
fitting.
2. Slide new O-ring over pointed end of
installer pin, onto the pin’s shank (see
Figure 34).
3. Insert pointed end of installer pin into open
end of installer sleeve until O-ring stops
against end of tool.
4. Position concave end of installer pin against
end of water/air port fitting (see Figure 35).
5. Push installer sleeve inward, until new O-
WARNING:
Avoid looking directly into optional curing light.
Protect patient with darkened eyewear when
using light probe near patient's eyes.
OPTIONAL EQUIPMENT STERILIZATION AND MAINTENANCE -
Cont'd:
O-RING
INSTALLER
SLEEVE
FIG. 35
INSTALLER
PIN
O-RING
INSTALLER
PIN
FIG. 34
O-RING
FIG. 36
WATER/AIR
FITTING
Fig. 33
O-Rings
(3)
LED