18.

OPTIONAL MOTOR LED LENS CLEANING:

The lens of the LED light on the optional electric

motor (Fig. 33) is soft and can be damaged. It

should not be exposed to dust and debris.

Excessive dust and debris may cause a drastic

decrease in optical output. In the event that the

light requires cleaning, first try a gentle

swabbing, using a lint-free swab. If needed,

use a lint-free swab and isopropyl alcohol to

gently remove dirt from the lens. 

Do not use

other solvents 

as they may adversely react

with the LED assembly.

OPTIONAL ULTRASONIC SCALER

:

(

Refer to ultrasonic scaler instructions for

use, supplied separately.)

The scaler handpiece cover and scaler tips are

fully autoclavable. Disinfect and clean the cover

and tips before autoclaving. Autoclave at a

maximum temperature of 135º C (275º F) for

10 minutes or 120º C (248º F) for 20 minutes.
Wipe off the scaler handpiece and its silicone

hose with a soft cloth. Use a 45% isopropyl and

detergent solution. 

DO NOT IMMERSE

the

handpiece in any fluid or spray any fluid directly

on the handpiece. 

OPTIONAL CURING LIGHT

:

(

Refer to curing light instructions for use,

Sunlite Lazer model, supplied separately.

Sunlite Lazer is a product of Kinetic Instruments, Inc.

)

The curing probe is fully autoclavable. Detach

probe by pressing quick disconnect button on

side of handle. Clean probe with disinfectant

then autoclave at 135°C (275°F) for 20 minutes

minimum. Clean and sterilize probe between

each patient use. 

NOTE:

The lamp module

(the section that seats into the tubing

connector) is NOT autoclavable.

OPTIONAL ELECTRIC MOTOR/CORD

RECEPTACLE O-RINGS

The O-rings for the three water/air ports in the

optional electric motor/cord receptacle should

be replaced if damaged or worn. Use the

provided O-ring installer pin and sleeve to

replace the O-rings:

1. Remove old O-ring from water or air port

fitting.

2. Slide new O-ring over pointed end of

installer pin, onto the pin’s shank (see

Figure 34).

3. Insert pointed end of installer pin into open

end of installer sleeve until O-ring stops

against end of tool. 

4. Position concave end of installer pin against

end of water/air port fitting (see Figure 35).

5. Push installer sleeve inward, until new      O-

WARNING:

Avoid looking directly into optional curing light.

Protect patient with darkened eyewear when

using light probe near patient's eyes.

OPTIONAL EQUIPMENT STERILIZATION AND MAINTENANCE - 

Cont'd:

O-RING

INSTALLER

SLEEVE

FIG. 35 

INSTALLER

PIN

O-RING

INSTALLER

PIN

FIG. 34 

O-RING

FIG. 36 

WATER/AIR

FITTING

Fig. 33

O-Rings

(3)

LED

Summary of Contents for AMC-20

Page 1: ...AMC 20 Mobile Dental System OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Page 2: ...e Note For reliable grounding connect to receptacle marked Hospital Grade To prevent injury to people and damage to property please heed relevant warnings and remarks They are marked as follows WARNIN...

Page 3: ...n E Type Endodontics Handpiece STANDARD FEATURES Automatic handpiece controls for two pneumatic handpieces Each pneumatic handpiece provides individually controlled water and coolant air 3 way air wat...

Page 4: ...ons concerning disposal of amalgam waste WARNING Never operate unit or work on patients when the top cover is open WARNING Do not use this device for dental implant procedures CAUTION Do not run salin...

Page 5: ...color coded to identify different fluids e g sterile water or distilled water 6 Place the foot control on the floor 7 Attach power cord to the back of the unit Fig 5 and plug into a grounded electric...

Page 6: ...tor into the amalgam manifold Tighten the retaining screw located under the top cover Fig 10 NOTE Before installing the separator into the manifold remove the two caps from the inlet and outlet ports...

Page 7: ...ESSURE RELEASE TOGGLE BOTTLE SELECTOR TOGGLE AUXILIARY WATER CONNECTION AUXILIARY AIR CONNECTION OPTIONAL SCALER COOLANT VALVE OPTIONAL HANDPIECE SCALER CURING LIGHT HOLDERS ELECTRIC MOTOR CONNECTOR H...

Page 8: ...10V OUTLET AIR TANK PURGE TOGGLE WASTE COMPARTMENT ACCESS DOOR OPTIONAL 220V OUTLET QUICK DIS CONNECT FOR WASTE HOSE 1 AMP CIRCUIT BREAKERS 2 TANK DRAIN OPTIONAL 10 AMP CIRCUIT BREAKERS 2 FOR 220V 20...

Page 9: ...PASS FILTER FILTER REGULATOR OPTIONAL FIBER OPTIC CURING LIGHT MODULE SYSTEM PRESSURE GAUGE PRESSURE REGULATOR CENTRAL VACUUM CANNISTER W AMALGAM SOLIDS TRAP INSIDE OPTIONAL OPTIONAL OPTIONAL OPTIONAL...

Page 10: ...oved from its holder and the foot control is pressed The amount of air pressure supplied to the pneumatic handpieces may be adjusted by turning the flow control screws on the control block OPERATION F...

Page 11: ...bout 5 seconds after every patient and about 20 seconds at the beginning of each day To flush a handpiece remove it from its holder and point the spray nozzle away from you i e into a basin Flip the f...

Page 12: ...atio s from the five available options Refer to page 12 for complete instructions on Customizing Ratios 3 Press the SPEED Up Down buttons to select the desired operating speed for the handpiece being...

Page 13: ...de press the Torque Up Down buttons to select the desired illumination in 10 increments from 10 to 100 The handpiece LED will automatically turn on and change intensity 4 Press the TORQUE Up Down butt...

Page 14: ...ttings a beep will sound 4 Press the foot pedal to activate the motor and begin operation as the adjustments are made The display icon will blink as the adjustments are made To exit the adjustment mod...

Page 15: ...100 On FACTORY DEFAULTS The optional Electric Motor Module will retain the factory default settings Fig 25 in memory until changed by the user To recall the original factory defaults press and hold th...

Page 16: ...ly electric scaler system that provides adjustable ultrasonic levels and a water coolant system The scaler includes 3 instrument tips The 37 instrument is specially developed for subgingival scaling f...

Page 17: ...alve and control valve are provided to turn coolant to the handpiece and motor On and Off and to control the flow Turn the coolant air control valve clockwise until it seats softly Begin turning count...

Page 18: ...t the water lines weekly Prepare a 1 10 bleach solution 1 part household bleach to 9 parts water Purge all water lines see page 17 Purging Water From the System Fill water bottles with bleach solution...

Page 19: ...ll water is purged from the lines with air Finally hold the water syringe over the basin and hold down the water button until all water is purged from the line with air GENERAL CLEANING The external s...

Page 20: ...e Lazer is a product of Kinetic Instruments Inc The curing probe is fully autoclavable Detach probe by pressing quick disconnect button on side of handle Clean probe with disinfectant then autoclave a...

Page 21: ...anchor eye bolts on the base as shown in Figure 38 7 Place the foot control pedal under the cart as shown in Figure 39 8 Wrap the cart with palletizer stretch wrap Fig 40 9 Insert the two end panels a...

Page 22: ...protective foam packing blocks on the front and top of the Cart Ensure the Cart s dental instrument holders are aligned with the cutouts on the top packing block carefully rotate the Cart s pivoting...

Page 23: ...ressure Insufficient vacuum Insufficient handpiece operation No water to handpiece No coolant air to handpiece Unit is turning On and Off Correction Check system power connection Check voltage selecto...

Page 24: ...w High Vol open 42 5 liters min 3 8 cm Hg Simultaneous Vacuum High 4 Hg 5 1 SCFM 144 4 liters min 10 2 cm Hg Low 1 5 Hg 1 4 SCFM 39 6 liters min 3 8 cm Hg Vacuum Compressor Pump 115 PSI 7 92 bar 2 HP...

Page 25: ...e Water Coolant Control Air Coolant ON OFF Toggle OR Flush Toggle Bottle Pressure Release Toggle Bottle Select Toggle Footswitch On Off Switch Mains Temperature Limitation Manufacturer Humidity Limita...

Page 26: ...LGAM SOLIDS TRAP OPEN TYPE AA 29 1 AMALGAM SOLIDS TRAP OPEN TYPE AA 29 100 100 AMALGAM SOLIDS TRAP CLOSED TYPE AA 290VS 1 AMALGAM SOLIDS TRAP CLOSED TYPE AA 290VS 100 100 VALVE DELUXE UNIVERSL AUTOCLA...

Page 27: ...of such action In the event of alleged defect under warranty the purchaser is to notify Aseptico s Customer Service Department promptly Customer Service will provide instructions usually directing th...

Page 28: ...P N 420869 Rev M ECO 14212 06 2017 PRINTED IN THE U S A P O Box 1548 Woodinville WA 98072 8333 216th Street SE Woodinville WA 98072 425 487 3157 800 426 5913 www aseptico com info aseptico com...

Reviews: