ASI Dental Specialties
2 – OPERATING ENVIRONMENT & CLASSIFICATION OF EQUIPMENT
www.asisupport.com
Page 3
65-0048 Rev-B (04/01/21)
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
This equipment
is intended for use in the electromagnetic environment specified below. The customer or the end user of this
equipment should assure it is used only in such an environment.
Immunity Test
EN/IEC 60601
Test Level
Compliance Level
Intended Electromagnetic
Environment
Electromagnetic Discharge (ESD)
EN/IEC 61000-4-2
± 8kV contact
± 2kV, ± 2kV, ± 8kV, ±
15kV air
± 8kV contact
± 2kV, ± 2kV, ± 8kV, ±
15kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast transient/burst
EN/IEC 61000-4-4
± 2kV for power supply
lines
± 1kV for input/output
lines
± 2kV for power supply
lines
± 1kV for input/output
lines
Mains power quality should be that of a
typical commercial or hospital environment.
Surge
EN/IEC 61000-4-5
± 1kV differential mode
(line-line)
± 2kV common mode
(line-earth)
± 1kV differential
mode (line-line)
± 2kV common mode
(line-earth)
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips, short interruptions
and voltage variations on power
supply input lines
EN/IEC 61000-4-11
0 % UT; 0,5 cycle at 0°,
45°, 90°, 135°, 180°, 225°,
270° and 315°
0 % UT; 1 cycle and 70 %
UT; 25 cycles Single
phase: at 0°
0 % UT; 250 cycle
0 % UT; 0,5 cycle at 0°,
45°, 90°, 135°, 180°,
225°, 270° and 315°
0 % UT; 1 cycle and 70
% UT; 25 cycles Single
phase: at 0°
0 % UT; 250 cycle
Mains power quality should be that of a
typical commercial or hospital environment. If
the user of this equipment requires continued
operation during power mains interruptions,
it is recommended that this equipment be
powered from an uninterruptible power
supply or a battery.
Power frequency (50/60Hz)
magnetic field
EN/IEC 61000-4-8
3A/m
3A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE:
UT is the a.c.
mains
voltage prior to application of the test level.