background image

ASI Dental Specialties 

2 – OPERATING ENVIRONMENT & CLASSIFICATION OF EQUIPMENT 

 

 

www.asisupport.com  

Page 4 

65-0048 Rev-B (04/01/21)  

Guidance and Manufacturer’s Declaration – Electromagnetic Immunity 

This equipment is intended for use in the electromagnetic environment specified below. The customer or the end user of this 

equipment

 

should assure it is used in such an environment. 

Immunity Test 

EN/IEC 60601  

Test Level 

Compliance 

Level 

Intended Electromagnetic Environment 

Conducted RF  

EN/IEC 61000-4-6 

Radiated RF 

EN/IEC 61000-4-3 

3Vrms between 

0.15MHz to 80MHz 

6V/m in ISM bands 

between 0.15MHz 

and 80MHz 

80MHz to 2.7GHz 

3Vrms between 

0.15MHz to 80MHz 

6V/m in ISM bands 

between 0.15MHz 

and 80MHz 

80MHz to 2.7GHz 

Portable and mobile RF communications equipment should be 

used no closer to any part of this equipment, including cables, 

than the recommended separation distance calculated from 

the equation applicable to the frequency of the transmitter. 

Recommended separation distance 

d = 1.2√P 

d = 1.2√P      80MHz to 800 MHz 

d = 2.3√P      800MHz to 2.5GHz 

where P is the maximum output power rating of the 

transmitter in watts (W) according to the transmitter 

manufacturer and d is the recommended minimum separation 

distance in meters (m). 

Field strengths from fixed RF transmitters, as determined by 

an electromagnetic site survey (a), should be less than the 

compliance level in each frequency range (b). 

Interference may occur in the vicinity of equipment marked 

with the following symbol: 

 

Proximity fields me 

the minimum 

requirement of  

      

Test Frequency 

(MHz) 

Band 

(MHz) 

Immunity Test Level 

(V/m) 

385 

380-390 

27 

450 

430-470 

28 

710 

704-787 

745 

780 

810 

800-960 

28 

870 

930 

1720 

1700-1990 

28 

1845 

1970 

2450 

2400-2570 

28 

5240 

5100-5800 

5500 

5785 

 

NOTE 1: 

 At 80MHz and 800MHz, the higher frequency range applies. 

NOTE 2: 

 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from objects, structures 

and people. 

a  Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and 

FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF 

transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which this equipment is used 

exceeds the applicable RF compliance level above, this equipment should be observed to verify normal operation. If abnormal performance is observed, 

additional measures may be necessary, such as re-orienting or relocating this equipment. 

b  Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m. 

 

 

 

Summary of Contents for 90-2115

Page 1: ...Models 90 2115 90 2125 This guide provides important steps for setup use and maintenance of the ASI delivery system It includes important details which must be followed to ensure safety and proper us...

Page 2: ...PRODUCT DESCRIPTION SPECIFICATIONS WARRANTY Intended Use Product Description Theory of Operation Dimensions Specifications Technical Data Warranty Information 4 UNPACKING INSTALLATION 5 START UP OPERA...

Page 3: ...Section 1 Contents Symbols Safety Precautions www asisupport com 65 0047 Rev A 11 12 20 SYMBOLS DESCRIPTION OF DANGER LEVELS GENERAL SAFETY PRECAUTIONS IDENTIFICATION OF SYMBOLS 65 OS01 Rev A 11 12 20...

Page 4: ...pressure before attaching or removing air lines or accessories or servicing this unit All air lines should be periodically inspected and replaced if worn or damaged If an outside compressed air suppl...

Page 5: ...ly with applicable regulation regarding EMC and shall be installed and put into service according to the EMC information stated below WARNING Use of portable phones or other portable or mobile radio f...

Page 6: ...ith UL 61010 1 2601 1 and under mutual recognition agreement with CAN CSA C22 2 No 601 1 Protective earth ground Functional earth ground Attention consult accompanying documents No user serviceable pa...

Page 7: ...n 2 Contents Operating Environment Classification of Equipment www asisupport com 65 0048 Rev B 04 01 21 CLASSIFICATION OF EQUIPMENT EN 60601 1 ENVIRONMENTAL SPECIFICATIONS ELECTROMAGNETIC COMPATIBILI...

Page 8: ...g where the intensity of electromagnetic disturbances is high ELECTROMAGNETIC COMPATIBILITY EMC Although this dental equipment conforms to the intent of the 2004 108 EC EMC Directive all medical equip...

Page 9: ...pecified in this manual or referenced documents may result in increased electromagnetic emissions from this equipment or decreased electromagnetic immunity of this equipment Guidance and Manufacturer...

Page 10: ...or hospital environment Surge EN IEC 61000 4 5 1kV differential mode line line 2kV common mode line earth 1kV differential mode line line 2kV common mode line earth Mains power quality should be that...

Page 11: ...RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the...

Page 12: ...pment Rated maximum output power of transmitter in watts W Separation distance according to frequency of transmitter in meters m 150kHz to 80MHz d 1 2 P 80MHz to 800MHz d 1 2 P 800MHz to 2 5GHz d 2 3...

Page 13: ...T DENTAL DELIVERY SYSTEMS PRODUCT DESCRIPTION MODELS AVAILABLE THEORY OF OPERATION OVERVIEW OF COMPONENTS AND HOW THEY WORK Handpiece Control and Operation Tandem Water Disinfection System Air Compres...

Page 14: ...and accessories Use of the system is intended only to support dental treatment and hygiene procedures and operation of dental instruments by or under the supervision of a licensed dentist All model co...

Page 15: ...CATIONS AND WARRANTY www asisupport com Page 2 65 0104 Rev A 04 01 21 OVERVIEW OF COMPONENTS AND HOW THEY WORK Handpiece Control and Operation Tandem Water System Compressed Air System o Air Compresso...

Page 16: ...nt and side of the unit Electric instruments such as the ultrasonic handpiece and electric motor are controlled via the same control block However the compressed air that flows from the block is route...

Page 17: ...y increases the volume of water available but also saves costs by only requiring the use of one water line cartridge The system works by allowing water in the second bottle to flow into the first wher...

Page 18: ...ces that enable a smooth running and quiet design The portable dental air compressor is controlled by an air pressure switch that senses when the pressure drops in the air tank to activate the motor a...

Page 19: ...t the end of the day The system separates the air stream from the liquid and provides a sediment bowl that can be removed Built into the top of the holding tank is a float switch to prevent over filli...

Page 20: ...ork done without first obtaining consent from ASI Specific Warranty Modifications The following items and or occurrences are not covered under the warranty Accidental Damage Theft including but not li...

Page 21: ...I s supplemental warranty does not extend the original term of the manufacturer s warranty does not include items not covered under the manufacturer s warranty or warranty problems caused by or result...

Page 22: ......

Page 23: ......

Page 24: ......

Page 25: ......

Page 26: ......

Page 27: ......

Page 28: ......

Page 29: ......

Page 30: ......

Page 31: ......

Page 32: ......

Page 33: ...Section 4 Contents Unpacking and Installation www asisupport com 65 0114 Rev A 02 08 21 UNPACKING ELECTRICAL INSTALLATION...

Page 34: ...ed to a central air supply appropriate electrical supply and optional vacuum and water supply Connection of umbilical lines from the cart to an appropriate junction box are the responsibility of the u...

Page 35: ...er Syringe Handpieces Fiber Optics HANDPIECE OPERATION Fiber Optics Operation Verify Handpiece Operation Handpieces Holders Coolant Operation Adjusting Coolant Spray Adjusting Chip Air Flow Oil Mist R...

Page 36: ...tch for supplying air pressure to the bottles Flip the switch to the OFF position to turn off the air supply to the bottles Slowly unscrew the bottles to allow them to depressurize Clean as appropriat...

Page 37: ...p assuring proper insertion and that the locking mechanism is engaged Handpieces Verify Handpiece Placements Ensure all handpieces are placed into their correct holders and that the corresponding tubi...

Page 38: ...pray allows water to flow up to high speed handpieces and to be sprayed onto the bur 5 Blank Hole This hole is non operative but is intended to still allow the use of a fiber optic handpiece in the st...

Page 39: ...Place the handpiece onto the coupler and ensure that the tubes from the handpiece align in the connector and that the gasket on the back of the handpiece is seated against the connector Then firmly ti...

Page 40: ...other instruments Water can also be sprayed from instruments to ensure it is working correctly Handpieces Holders The handpieces are automatically activated for use with the foot control whenever the...

Page 41: ...also be desired to shut down chip air flow for some treatment procedures In this case rotate the knob counter clockwise until it stops Oil Mist Recovery Jar Air from an air driven handpiece is exhaus...

Page 42: ...never it is not in use The strength of the vacuum pull to the oral evacuation system is regulated by the vacuum relief valve The relief valve on the system is pre set to approximately 12 Hg which is i...

Page 43: ...Exterior Cleaning and Worktop Worktops and Control Panel Overlays Painted Work Surfaces and Handpiece Holders Wiping Down Disinfecting Handpiece Tubings Disinfection of Clean Water System Handpiece St...

Page 44: ...n the suction canister The canister holds up to 2 liters of solution When the canister is full an overflow shut off device will disable the vacuum pump until purged IMPORTANT Do not fill the canister...

Page 45: ...n OFF the purge pump switch and rewind the purge hose 8 After the diluted solution is purged out from Step 2 above suction up 2 ounces of pure concentrated Keep Clear solution or place an atomizer cap...

Page 46: ...dental unit utilizes a container to assist in capturing excess handpiece lubricant This Oil Mist Recovery Jar is located under the left side of the system behind the holder bar Exhaust lines of the pn...

Page 47: ...eaning and Worktop The exterior of the delivery unit should be thoroughly cleaned and disinfected after use each day to maintain optimal performance All components should be maintained in proper worki...

Page 48: ...rtridge will no longer provide disinfection for the system and may plug the unit and obstruct the water flow Do not use the DentaPure DP365 if using a shock treatment to disinfect lines as it will dam...

Page 49: ...pieces 3 Remove the HVE or saliva ejector valve micro evacuator body from the tubing by pulling it away from the tubing to disconnect from the barb end that remains in the tubing 4 Remove the control...

Page 50: ...Empty the water containers and purge handpiece coolant lines and the water syringe See Operation Instructions Run air through the handpieces for an additional three minutes to air dry 2 Purge the was...

Page 51: ...nd tie to secure Remove the lid from the trunk and set the main part of the trunk on end Lift out the caster ramp and remove the foam insert from the top Carefully roll and push the cart into the trun...

Page 52: ...ssure to the Handpiece Air or Water Coming out of Handpieces When Hung Up Water Filling Up Oil Mist Recovery Jar Water Not Flowing From Handpieces or Air Water Syringe Water Flows from Air Water Syrin...

Page 53: ...at handpiece has been opened Ensure that the exhaust air tubes and oil mist recovery jar are allowing the air to exhaust freely Verify that there is no kinked exhaust tubing to the oil mist recovery j...

Page 54: ...the control block and verify that the ridged side of the diaphragm puck is facing down Water Not Flowing From Handpieces or Air Water Syringe Check the water bottles and ensure they are full switched...

Page 55: ...ation If water drips from several handpieces or if water slowly builds and drips out the water relay may not be working or has a small piece of debris in it Replace the water relay valve o Replace wit...

Page 56: ...ing Grease Use only silicone lubricant when lubricating instrumentation O rings Petroleum products will cause permanent damage to the O rings If air is blowing out of foot control the diaphragm in the...

Page 57: ...ervicing Preventative Maintenance www asisupport com 65 0059 Rev C 04 27 21 CUSTOMER SUPPORT REPLACEMENT PARTS List of Replacement Parts Kits for this System SERVICE PREVENTATIVE MAINTENANCE DAILY 1 Y...

Page 58: ...115V PN 95 0416 Air Compressor Manifold DV PN 95 0417 Compressor and Vacuum Rebuild Kit PN 95 0378 Check Valve Repair Kit PN 95 0108 Vacuum Pump 115V PN 95 0167 Vacuum Pump DV PN 95 0167D Capacitor fo...

Page 59: ...ERVICE Please follow the following steps for service of your equipment Start with Section 7 Troubleshooting This section is designed to help narrow down most common problems Visit ASIsupport com Tech...

Page 60: ...oggle PN 95 0242 if needed X CLOSED WATER SYSTEM Daily 1 Year 5 Years Turn air pressure toggle to ON Verify closed water system gauge is operational Check for cracks or air leaks Replace gauge PN 95 0...

Page 61: ...ct all vacuum tubing and fittings for leaks or damage Replace if needed X Replace felt filter muffler PN 95 0093 X Install rebuild kit for vacuum pump every 5 10 years if needed Serial Number specific...

Page 62: ...inge Closed Water System Instrument Component Pan View Air Compressor Complete Assembly Diagram Air Tank Dual Bracketed Complete Assembly Vacuum Pump Complete Assembly Diagram Sediment Separator and H...

Page 63: ...ASI Dental Specialties 9 BACK VIEW DIAGRAM TRITON www asisupport com Page 1 65 0062 Rev A 11 12 20 BACK VIEW TRITON 65 OS01 Rev A 11 12 20...

Page 64: ...ASI Dental Specialties 9 CHASSIS INTERIOR DIAGRAM TRITON www asisupport com Page 1 65 0063 Rev A 11 12 20 CHASSIS INTERIOR VIEW TRITON...

Page 65: ...ASI Dental Specialties DELIVERY SYSTEM TUBING SCHEMATIC www asisupport com Page 1 65 0064 Rev A 11 12 20 DELIVERY SYSTEM TUBING 65 OS01 Rev A 11 12 20...

Page 66: ...tal Specialties 9 CONTROL BLOCK DIAGRAM www asisupport com Page 1 65 0065 Rev B 02 08 21 5 POSITION CONTROL BLOCK FRONT OUTPUT CONNECTIONS TO HANDPIECE 5 POSITION SINGLE CONTROL BLOCK SIDE INPUT CONNE...

Page 67: ...ASI Dental Specialties 9 FOOT CONTROL DIAGRAM www asisupport com Page 1 65 0069 Rev A 11 12 20 FOOT CONTROL 65 OS01 Rev A 11 12 20...

Page 68: ...in Quick Clean Syringe Button 3a Air Only Syringe 3b Air Water Syringe 4 Tip Adapter Assembly 5 Large O Ring Tip Adapter 6 Spring Tip Adapter 7 O Ring Tip Adapter 8 Button Assembly 9 O Ring Button Ass...

Page 69: ...ASI Dental Specialties 9 CLOSED WATER SYSTEM DIAGRAM www asisupport com Page 1 65 0071 Rev A 11 12 20 CLOSED WATER SYSTEM 65 OS01 Rev A 11 12 20...

Page 70: ...le 1 PN95 0198 Power Supply Curing Light 1 Power Supply Kavo Motor 1 Controller Fiber Optic System 1 PN 95 0229 Power Controller Kavo Motor 1 Oil Mist Recovery Jar Handpiece Exhaust 1 PN 95 0258 Trans...

Page 71: ...ASI Dental Specialties 9 AIR COMPRESSOR DIAGRAM www asisupport com Page 1 65 0073 Rev A 11 12 20 AIR COMPRESSOR ASSEMBLIES Single Voltage PN 95 0197 Dual Voltage PN 95 0159D 65 OS01 Rev A 11 12 20...

Page 72: ...ASI Dental Specialties 9 AIR COMPRESSOR TANK ASSEMBLY DIAGRAM www asisupport com Page 1 65 0074 Rev A 11 12 20 AIR TANK ASSEMBLY 80 0192 65 OS01 Rev A 11 12 20...

Page 73: ...ASI Dental Specialties 9 VACUUM PUMP ASSEMBLY DIAGRAM www asisupport com Page 1 65 0075 Rev A 11 12 20 VACUUM PUMP ASSEMBLIES Single Voltage PN 95 0167 Dual Voltage PN 95 0167D 65 OS01 Rev A 11 12 20...

Page 74: ...ASI Dental Specialties 9 SEDIMENT SEPARATOR DIAGRAM www asisupport com Page 1 65 0076 Rev A 11 12 20 SEDIMENT SEPARATOR HOLDING TANK PN 95 0343 65 OS01 Rev A 11 12 20...

Page 75: ...ASI Dental Specialties 9 ELECTRICAL WIRING SCHEMATIC 115V www asisupport com Page 1 65 0077 Rev A 11 12 20 WIRING FOR CART CHASSIS 115V 65 OS01 Rev A 11 12 20...

Page 76: ...ASI Dental Specialties 9 ELECTRICAL WIRING SCHEMATIC 115V www asisupport com Page 2 65 0077 Rev A 11 12 20 WIRING FOR CART BASEPLATE 115V 65 OS01 Rev A 11 12 20...

Page 77: ...ASI Dental Specialties 9 PURGE PUMP ASSEMBLY DIAGRAM 115V www asisupport com Page 1 65 0078 Rev A 11 12 20 PURGE PUMP ASSEMBLY 115V PN 95 0404 65 OS01 Rev A 11 12 20...

Page 78: ...ASI Dental Specialties 9 PRESSURE SWITCH DIAGRAM 115V www asisupport com Page 1 65 0079 Rev A 11 12 20 PRESSURE SWITCH 115V 65 OS01 Rev A 11 12 20...

Page 79: ...ASI Dental Specialties 9 ELECTRICAL WIRING SCHEMATIC DV www asisupport com Page 1 65 0080 Rev A 11 12 20 WIRING FOR DUAL VOLTAGE CHASSIS 65 OS01 Rev A 11 12 20...

Page 80: ...ASI Dental Specialties 9 ELECTRICAL WIRING SCHEMATIC DV www asisupport com Page 2 65 0080 Rev A 11 12 20 WIRING FOR DUAL VOLTAGE BASEPLATE 65 OS01 Rev A 11 12 20...

Page 81: ...ASI Dental Specialties 9 PURGE PUMP ASSEMBLY DIAGRAM DV www asisupport com Page 1 65 0081 Rev A 11 12 20 PURGE PUMP ASSEMBLY PN 95 0293 65 OS01 Rev A 11 12 20...

Page 82: ...ASI Dental Specialties 9 PRESSURE SWITCH ASSEMBLY DIAGRAM DV www asisupport com Page 1 65 0083 Rev A 11 12 20 PRESSURE SWITCH ASSEMBLY DUAL VOLTAGE 65 OS01 Rev A 11 12 20...

Reviews: