ASI Dental Specialties
5 – START UP, OPERATION & SHUT DOWN
www.asisupport.com
Page 6
65-0067 Rev-A (11/12/20)
To adjust coolant spray, press the foot control and while the handpiece
is operating, locate the adjustment knob on the unit on the side or
front. Turn the toggle switch for coolant spray to ON, then turn the
adjustment knob (counter-clockwise to decrease, or clockwise to
increase) until the desired level of spray is achieved (see figure).
For multiple instruments that provide coolant, individual
adjustments are provided and are located in order from left
to right, and are labeled accordingly.
Adjusting Chip Air Flow
The Chip Air Control is located on the front left side of the unit and is labeled. To adjust, rotate clockwise to
increase and counter-clockwise to decrease the flow. Depending on the brand of handpiece, chip air may be
used to create the desired atomization of the coolant spray. It may also be desired to shut down chip air flow for
some treatment procedures. In this case, rotate the knob counter clockwise until it stops.
Oil Mist Recovery Jar
Air from an air driven handpiece is exhausted through the tubing and exits
through the oil mist recovery jar. The oil mist recovery jar helps to trap
droplets of excess oil used in handpiece lubrication.
OTHER INSTALLED INSTRUMENTS
Please refer to Section 10 – References or Supplemental Information, for separate instructions for each
individual instrument that may be installed as an accessory in the system.
ORAL EVACUATION SYSTEM
To use an evacuation handpiece, remove the valve from the holder and insert the desired tip into the valve. The
saliva ejector valve accommodates standard disposable saliva ejector tips and the HVE valve accommodates
standard disposable 7/8” tips.
To ensure maximum suction flow, periodically check and replace the solids trap screen located on the front of
the system. Remove the cover to the solids trap, lift out the old screen, insert the new screen and replace the
cover to the solids trap.
65-OS01 Rev-A | 11/12/20