SECTION 2 SYSTEM SETUP AND PREPARATION FOR USE
______________________________________________________________________
2-1
SECTION 2
2
SYSTEM SETUP AND PREPARATION
FOR USE
____________________________________________________
INTRODUCTION
This section provides setup instructions for the Aspect BIS VIEW Monitor, BISx, and
accessories. It includes:
•
Setup checklist
•
Proper environment
•
Required equipment and supplies
•
Cable connections
•
Start and shutdown procedures
•
Initial menu settings
2.1
BIS VIEW Monitor Setup and Checkout
1.
Open packages and inspect for all components:
•
Monitor (P/N 185-0205)
•
Power cord
•
Pole clamp
•
BISx (P/N 185-0145-AMS)
•
PIC (Patient interface cable, connects BISx to patient)
Sensors are sold separately. For a list of available sensors please contact Aspect Medical
Systems, Inc. or your local distributor.
2.
Connect power cable to monitor, plug power plug into appropriate wall outlet.
•
Verify that light to right of ON/Standby button is yellow.
3.
Start up monitor by pressing the ON/Standby button (lower right corner).
•
Verify that light to right of ON/Standby button is green.
•
Verify all self-tests complete successfully. (Do not connect or disconnect equipment,
or press keys until the monitor has completed its tests.)
•
Verify next screen says “Connect BISx.”
4.
Connect BISx with PIC to monitor.
•
Verify that screen says “Connect sensor or cable.”
5.
Connect PIC and sensor.
•
Verify SENSOR CHECK begins.
6.
Disconnect power cord from monitor.