SECTION 6 DIAGNOSTICS AND TROUBLESHOOTING
______________________________________________________________________
6-4
6.2.2
Diagnostic Codes
If you are experiencing problems with the BIS VIEW system, Aspect’s Technical Service
may direct you to turn the Diagnostic Codes ON so that numeric diagnostic values will be
displayed in the Message Region of the screen.
To turn ON the Diagnostics Codes:
1.
Press
[MENU]
. The Main Menu displays.
2.
Use the
[
▲
]
or
[
▼
]
key to highlight
[Diagnostics]
. Press
[SELECT]
. The
Diagnostics Menu displays.
3.
Highlight
[Diagnostic Codes]
. Press
[SELECT]
.
4.
Highlight
[On]
. Press
[SELECT]
.
5.
Press
[HOME]
to exit.
6.2.3
System Configuration Information
This selection allows the user to view the current configuration of the system, including the
monitor, BISx, and sensor. It also shows the revision of each language installed in the
monitor.
To view System Configuration Information:
1.
Press
[MENU]
. The Main Menu displays.
2.
Use the
[
▲
]
or
[
▼
]
key to highlight
[Diagnostics]
. Press
[SELECT]
. The
Diagnostics Menu displays.
3.
Highlight
[System Configuration]
. Press
[SELECT]
.
4.
Use the
[MORE]
key to view the next screen, or the
[HOME]
key to exit.
6.2.4
Advanced Diagnostics
The Advanced Diagnostics Menu contains system information and test procedures for the
monitor and the BISx. Contact Aspect Medical Systems, Inc. Technical Service Department
for assistance on using these functions. Contact information is located on the back page of
this manual.