Operation manual
Ergometer CRG 200 v.5xx
IU-AT-77-C_I_ENG
ASPEL S.A.
19.05.2020
Strona 15 z 31
Safety of use:
EN 60601-1
Electromagnetic compatibility:
EN 60601-2
Medical device class:
IIa (rule 10)
Type of anti-electric shock
protection (EN 60601-1):
Class II
Classification of training equipment
(EN 957-1):
Class SA
Equipment class and group
acc. to CISPR-11:
Class A, group 1
IP protection class:
IP X0
ECG signals:
12 standard leads (diagnostic mode),
2 leads (training mode)
Sensitivity:
2,5/5/10/20 mm/mV ± 5%
Recording speed:
25/50/100 mm/s ± 5%
Frequency range:
0,05 ÷ 150 Hz
HR measuring range:
15 - 240 bpm
HR measuring error:
± 2 %
HR resolution:
1 bpm
Digital filters:
25 Hz, 35 Hz, 50 Hz, anti-drift filter
Monitoring range of ST section:
± 3 mm (10mm/mV)
ST measuring error:
± 0,2 mm (10mm/mV)
ST resolution:
0,1 mm (10mm/mV)
Applied part (EN 60601-1):
defibrillation-resistant type CF
Input circuit is secured against defibrillating impulse. After the impulse, the ECG line
will appear after no longer than 10 seconds.
2.6. Equipment
•
AC/DC power supply CRG 200-M24.
•
Power cable.
•
Data cable.
•
KEKG 51 v.103 patient cable (for an exercise test).
•
KEKG 52 v.105 patient cable (for cardiological rehabilitation).
•
Disposable ECG leads (50 pcs).
•
Abrasive paste.
•
Warranty card.
ECG parameters: