Operation Manual Touchscreen
8. STANDBY
To put the drying cabinet in standby mode, press the [Standby] button via [Menu] and then [Activate
standby].
While the standby mode is activated, the backlighting of the display and the drying unit are
deactivated. The fan of the optional heater can only be stopped by a complete power disconnect.
During regeneration, standby mode is not available.
ATTENTION: Always disconnect the power supply during maintenance!
9. DATA LOGGING
The values for humidity and temperature are logged every minute. In addition, relevant events are
logged as: cabinet opening, alarm releases, alarm acknowledgments, ...)
One file is automatically created per day in the "SAMP02" folder. The file name is named after the
current date in the format: SAmmddy (example: 3.21.2017 SA03217).
The locking works only with a plugged USB storage medium!
The file can either be read directly from the USB stick or downloaded via FTP.
ftp://[IP address of display] in the web browser.
To read out the csv file, Excel program can be used by Windows. The free Totech Viewer offers the
ability to download the data graphically.
10. CALIBRATION
Totech recommends to check the calibration once a year or to recalibrate the system by replacing the
sensor.
To calibrate the sensor, we recommend the following two possibilities:
1.
Replacing the sensor by a factory-calibrated sensor. After detaching the screw-fastened retaining
ring, the plug-in sensor can simply be unplugged. The replacement sensor is mounted in reverse
order. Since all settings have been stored in the sensor, its replacement will serve to recalibrate the
dehumidifier's overall measuring. Additional measurement equipment or software is not required. After
replacing the sensor for the new calibrated sensor, the old sensor has to be returned to the factory.
You do not need any further measuring devices or software.
2.
Define deviations by means of a calibrated reference dew point meter (minimum accuracy
±0.2%RH or higher), and calibrate the sensor via the software's calibration function. This proceeding
requires data connection to PC and software (for detailed instructions, please refer to the Software
Manual of the manufacturer).
(see 6.2.1 Offset, Lock, Firmware).