Operation Manual Touchscreen
11.3 SELECT THE LOGO! BASE MODULE IP ADDRESS
To make sure the display is connected to the LOGO! Base module in the drying unit, the IP address of
the base module has to be specified in the display:
To do this, go to the [Setting] menu on the 3rd page and press [Offline].
Press [Peripheral]
→
[Device/PLC Settings]
→
[Siemens AG SIMATIC S7 Ethernet]
→
[Device]
→
“DEST IP ADDRESS”
⇒
Here you have to enter the IP address which you have given to the LOGO!
base module.
(NOTE: Depending on the display type and firmware, the display may vary slightly.)
12. Calibration
Totech recommends annual calibration inspection or the system's calibration by sensor replacement,
respectively. For the drying unit's calibration, we recommend the following two options:
1. Replacing the sensor by a factory-calibrated sensor. After detaching the screw-fastened
retaining ring, the plug-in sensor can simply be unplugged. The replacement sensor is
mounted in reverse order. Since all settings have been stored in the sensor, its replacement
will serve to calibrate the dehumidifier overall measuring. Additional measurement equipment
or software is not required.
2. Define deviations by means of a calibrated reference dew point meter (minimum accuracy
±0.2%RH or higher), and calibrate the sensor via the software's calibration function. This
proceeding requires data connection to PC and software (for detailed instructions, please refer
to the Software Manual of the manufacturer).
13. Maintenance
Ensure to generally disconnect the dehumidifier from the electrical power supply before opening or
disassembling the device.
Ensure all motors - in particular ventilators - are at standstill before reaching inside the device!
In case the dehumidifier was in operation, please allow a cooling time of at least 30 minutes before
disassembling the device!
General maintenance; Dehumidifier U-5000 has been designed for maintenance-free long-term
operation. A maintenance schedule does thus not exist.
Sensor replacement and calibration; please proceed as described under item 7.
Repair works are to be performed by qualified Totech service engineers or adequately trained
personnel only. Please contact your local Totech dealer, or Totech EU directly at the service portal
found on the website:
http://service.superdry-totech.com/