22
V
ision
R
ecorder
C
HANGING THE
T
EMPERATURE
C
ALIBRATION
T
o
change
the
temperature
calibration
:
1. D
isplay
the
CAL
b
root
menu
.
2. P
ress
. T
he
number
0.0
appears
.
3. P
ress
or
to
raise
or
lower
the
temperature
measurement
,
in
steps
of
0.1
°
. Y
ou
can
raise
or
lower
the
measurement
by
-2
°
to
2
°
in
one
action
.
4. W
hen
you
are
finished
,
press
.
CAL
b
appears
.
5. P
ress
to
save
your
changes
,
exit
configuration
,
and
display
the
current
temperature
.
E
RASING THE
L
OG
T
he
AT-VLT
saves
the
last
temperature
measurements
(
up
to
290
logs
). T
o
erase
the
log
:
1. D
isplay
the
L
o
9
r
root
menu
.
2. P
ress
. T
he
log
is
erased
,
and
AL
r
appears
.
3. T
he
AT-VLT
will
and
or
send
the
message
MEMORY RESET BY USER,
<
date
>, <
time
>
in
the
currently
configured
format
.
4. P
ress
to
exit
configuration
and
display
the
current
temperature
.
C
HANGING THE
S
AMPLING
R
ATE
T
he
AT-VLT
measures
temperatures
at
defined
intervals
. T
o
change
the
intervals
at
which
temperatures
are
measured
:
1. D
isplay
the
SAP
r
root
menu
.
2. P
ress
. T
he
current
sampling
rate
appears
. T
he
sampling
rate
can
be
any
value
between
1
and
90
minutes
.
3. P
ress
or
to
change
the
sampling
rate
. W
hen
you
are
finished
,
press
.
SAP
r
appears
.
4. P
ress
to
save
your
changes
,
exit
configuration
,
and
display
the
current
temperature
.