Manual for use of thermo-hygro-barometer ATAL ATC10 with external probe
Instrument is designed for measurement of temperature, relative humidity, atmospheric
pressure and pressure tendency. It enables to display the dew point temperature and atmospheric
pressure corrected to the sea level. Measured values are displayed on a dual line LCD display.
Temperature is measured by RTD sensor Ni1000/6180ppm. Instrument is equipped with an
external temperature - relative humidity probe. Instrument compares all measured values (except
pressure tendency) with two adjustable levels for each measured quantity. Breaking the level is
indicated by blinking the proper value on display and by audio indication (switchable). Instrument
is equipped with minimum and maximum memory and Hold function. Minimum and maximum
values and Hold value are possible to display on the LCD anytime.
Technical parameters:
Parameters of measurement:
Temperature: Range of measurement: internal sensor (INT) -10 to +60 °C
external sensor (EXT) -30 to +105 °C (temperature
from +85 to +105°C with the limitation accordingly to the graph below)
Resolution: 0.1 °C
Accuracy: ±0.4 °C
Humidity:
Range measurement: 0 to 100 %RH
Resolution: 0.1 %RH
Accuracy: ±2.5 % RH at the range of 5 to 95 %RH
Dew point temperature (calculated from temperature and humidity):
Range: -50 to +60 °C
Resolution: 0.1 °C
Accuracy: ±0.5 °C at the range of 30 to 95 % RH
Atmospheric pressure: Range of measurement: 800 to 1100hPa
Resolution: 0.1 hPa
Accuracy: ± 1hPa at ambient temperature of 23°C
Measuring interval and display reading refresh: approximately 0.7 s in FAST mode
0.7 to 5 s in dynamic mode
Interval of refresh of pressure tendency display: 0.5 h
Power: battery 9V or ac/dc adapter 12V with NiCd accumulator 9V
Typical life of Zinc-Chloride battery: 2 months
Typical life of Alkali-Mangan battery: 3 months
Average current consumption: 0.3 to 1.3 mA (depending on operation mode)
Operation conditions:
Ambient temperature range: -10 to +60 °C
Ambient relative humidity range: 5 to 95 %RH
Outer characteristics in accordance with EN 33-2000-3: normal environment with
characteristics AD1, AE1, AF1, AG1, AH1, AK1, AL1, AN1, AP1, AQ1, AR1,
AS1, BA1, BE1
Not allowed manipulation: it is not allowed to touch sensors under the cover to avoid
sensors damaging or to effect calibration.
The sensors (under the cover) should not be exposed to direct contact with water or other
liquids.
Storing conditions: temperature -10 to +60 °C relative humidity 5 to 95 %RH
Dimensions: 141 x 71 x 27 mm
Weight including battery: approximately 200 g