Chapter 3 –
Start up, adjust and measures
UM-23900B-U User
manual
ATEQ CDF60
Page 16/40
3.4. G
AS
This parameter allows calculating the flow measurement for a other gas, default gas:
air.
The avalaible parameters are for the following gas:
¾
Helium (He),
¾
Nitrogen (N
2
),
¾
Other (the parameter had to be indicated, consult us).
Enter into the parameters menu, by pressing and
old the
ENTER
during 3 seconds.
+ / - 4 0 . 0 0 c m 3 / m i n
A I R S T D . C O N D .
FLOW :
09.36
TEMP. :
25.8°C
ATM.P :
1010.5 hPa
The
PARAMETER
menu is displayed. By using the
UP
and
DOWN
key select
GAZ
parameter.
PARAMETER
FLOW UNIT.: cm3/min
STD.COND.: NO
>GAZ : AIR
Then press the
ENTER
key to modify the
parameter available the gas list is displayed. By
using the
UP
and
DOWN
key choose
the gas to be measured.
GAZ
>AIR
HELIUM
NITROGEN
It's possible to configure a different gas than those
available, for that select the
OTHER
option, then
configure the applied coefficient. For further
information, contact ATEQ company.
GAZ
HELIUM
NITROGEN
>OTHER
Validate with the
ENTER
key.
GAZ
>AIR
HELIUM
NITROGEN
To return to the measurement mode, press
CANCEL
.
+ / - 4 0 . 0 0 c m 3 / m i n
A I R S T D . C O N D .
FLOW :
09.36
TEMP. :
25.8°C
ATM.P :
1010.5 hPa
Summary of Contents for CDF60
Page 1: ...USER MANUAL ATEQ CDF60 Version 1 0 Reference UM 23900B U...
Page 2: ......
Page 4: ......
Page 6: ......
Page 8: ......
Page 10: ......
Page 20: ...Chapter 2 User Interfaces UM 23900B U User manual ATEQ CDF60 Page 10 40...
Page 40: ...Chapter 6 Error messages UM 23900B U User manual ATEQ CDF60 Page 30 40...
Page 46: ...Chapter 7 PC installation UM 23900B U User manual ATEQ CDF60 Page 36 40...
Page 50: ...Index UM 23900B U User manual ATEQ CDF60 Page 40 40...
Page 51: ......