Chapter 2 – 

User Interfaces 

 

UM-24100D-U User 

manual

 ATEQ CDP60 

Page 7/36 

 

Chapter 2 

 USER INTERFACES 

1. KEYS 

1.1. “S

TART 

/

 

I

NFORMATION

 KEY

 

KEY

 

FUNCTION

 

 

Switched off instrument

: this key switch on the instrument. 

Switched on instrument

: allows the displaying of:  

¾

 the program 

version

,  

¾

 the pressure 

range

,  

¾

 the 

battery

 level.  

¾

  Press and hold this key to 

switch of

 the instrument 

(more than 3 seconds). 

Press and hold the key to display the battery level.  

1.2. "C

ANCEL

"

 KEY

 

KEY

 

FUNCTION

 

 

Cancel he edition or the modification of the parameters. 

Return to the previous menu. 

1.3. “HOLD”

 KEY

 

KEY

 

FUNCTION

 

 

To freeze or unfreeze the measurement display.   

 

Summary of Contents for CDP60

Page 1: ...USER MANUAL ATEQ CDP60 Version 1 03 Reference UM 24100D U ...

Page 2: ......

Page 3: ...n UM 24100A U 25 2007 Second edition UM 24100B U 11 2008 Modification of the measurement characteristics in the preamble Third edition UM 24100C U 26 2008 Program evolution to the 1 03 version Fourth edition UM 24100D U 44 2009 Preamble modification of the principle scheme Chapter 5 pressure generator use ...

Page 4: ......

Page 5: ...l that the electrical main is earthed disconnect the electrical connections to the equipment before maintenance cut the air supply for any kinds of operation on the pneumatic assembly do not open the instrument when it is powered up avoid water spillage near of the instrument ATEQ is at your disposal for any further information concerning the use of the instrument under maximum safety conditions W...

Page 6: ......

Page 7: ...ments CALIBRATION This may be carried out on site or in our offices ATEQ is attached to the COFRAC and delivers a certificate following a calibration TRAINING COURSES In the framework of partnership with our customers ATEQ offers two types of training in order to optimise the usage and knowledge of our instruments They are aimed at different levels of technician method control training maintenance...

Page 8: ......

Page 9: ...ve you complete satisfaction during many long years of operation In order to maximise the life expectancy and reliability of your ATEQ instrument we recommend that you install this instrument on a secured workbench and advise you to consult this manual in order to familiarise yourself with the functions and capabilities of the instrument Our ATEQ After Sales Service centre can give you recommendat...

Page 10: ......

Page 11: ...nput 6 2 5 Pneumatic connector second input 6 Chapter 2 USER INTERFACES 1 KEYS 7 1 1 Start Information key 7 1 2 Cancel key 7 1 3 HOLD key 7 1 4 Enter key 8 1 5 Navigation keys 8 2 OTHERS ITEMS 9 2 1 LCD display 9 2 2 Lights 9 Chapter 3 START UP ADJUST AND MEASURES 1 STARTING THE ATEQ CDP60 11 2 MEASURES DISPLAY 11 3 PARAMETERS ADJUST 12 3 1 Filter 12 3 2 Pressure unit 13 3 3 Back light 14 3 4 Aut...

Page 12: ...22 1 5 Protection cover option 22 1 6 Hard case option 22 1 7 Pressure generator option 22 2 RECYCLING 23 3 SAFETY INFORMATIONS 24 Chapter 6 ERROR MESSAGES 1 ERROR MESSAGES 25 Chapter 7 PC INSTALLATION 1 PROGRAM AND DRIVERS INSTALLATION 27 1 1 USB connection 27 1 2 RS232 connection 31 1 3 RS232 cable diagram 31 Appendix ATEQ CDP60 1 TECHNICALS CHARACTERISTICS 33 2 PERSONNAL NOTES 34 Index 35 ...

Page 13: ...ION OF THE ATEQ CDP60 The CDP60 is a master used to adjust or calibrate all kinds of instruments measuring air pressure PRESSURE SENSOR ATMOSPHERE AUTO ZERO ELECTROVALVE ATEQ CDP60 The auto zero electrovalve is integrated into instruments with a full scale less than 5 bars ...

Page 14: ...500 hPa 0 20 RV 1 hPa 4000 hPa 0 9 RV 0 01 FS C 0 1 hPa 1000 5000 hPa 0 20 RV 5 hPa 7000 hPa 0 5 RV 0 01 FS C 1 hPa 0 1 2 MPa 0 20 RV 2 kPa 3 MPa 0 5 RV 0 01 FS C 10 hPa FS Full scale RV Read value Specifications in option adjusted in our metrology laboratory with COFRAC agreement Note the zero offset linked to the temperature or the position can be take over by an auto zero It is however advised ...

Page 15: ...ing a portable case The instrument has a pneumatic quick connector Staublï kind for the connection to the pressure to be measured The instrument has three electrics connectors 1 connector for the battery charge the instrument run with lithium ion batteries under 12 V DC voltages and it supplies with a specific supply transformer to charge the batteries 1 RS232 connector 1 USB connector Display Key...

Page 16: ...ector 2 3 RS232 CONNECTOR Allows the connection to a PC or a printer with a RS232 connection RJ45 type connector For further information see chapter 7 paragraph 1 2 RS232 connection 2 4 PNEUMATIC CONNECTOR FIRST INPUT Allows the connection to the pressure to be measured quick connector Staublï RBE03 female type 2 5 PNEUMATIC CONNECTOR SECOND INPUT In case of low pressure differential sensor calibr...

Page 17: ...hed on instrument allows the displaying of the program version the pressure range the battery level Press and hold this key to switch of the instrument more than 3 seconds Press and hold the key to display the battery level 1 2 CANCEL KEY KEY FUNCTION Cancel he edition or the modification of the parameters Return to the previous menu 1 3 HOLD KEY KEY FUNCTION To freeze or unfreeze the measurement ...

Page 18: ...cond allows acceding to the special cycles menu 2nd mode press and hold 3 seconds allows acceding to the parameters menu 1 5 NAVIGATION KEYS KEY FUNCTION Scroll up or increase numeric values Swing from the single Pressure or extended Pressure temperature Atmospheric pressure displays Scroll down or decrease numeric values Swing from the single Pressure or extended Pressure temperature Atmospheric ...

Page 19: ...nstrument state Low Bat Battery level when this is on the battery level becoming too low for reliable operation indicating that the battery must be recharged before further use Charge Charge light this light is glowing red while the battery is being charged When charging is finished the light turns off Tx Not used Note when the device is connected to a PC USB port the Charge light is on ...

Page 20: ...Chapter 2 User Interfaces UM 24100D U User manual ATEQ CDP60 Page 10 36 ...

Page 21: ...le superior than 5 bars the instrument asks to disconnect all pressure source from the instrument for making the auto zero When it s disconnect press the ENTER key Switch off pressure Enter If Ok The instrument starts the auto zero cycle AUTO ZERO At the end of the auto zero cycle the instrument indicates the current pressure measured and is ready for measurement 500 0 Pa 000 0 Note it s important...

Page 22: ...TER during 3 seconds 500 0 Pa 000 0 The PARAMETER menu is displayed By using the UP and DOWN key select FILTER parameter PARAMETER FILTER 0 5 s PRESS UNIT Pa BACK LIGHT 100 Then press the ENTER key to modify the parameter the cursor moves on the right PARAMETER FILTER 0 5 s PRESS UNIT Pa BACK LIGHT 100 By using the UP and DOWN key modify the parameter to the hoped value PARAMETER FILTER 1 0 s PRES...

Page 23: ...TER menu is displayed By using the UP and DOWN key select PRESS UNIT parameter PARAMETER FILTER 0 5 s PRESS UNIT Pa BACK LIGHT 100 Then press the ENTER key to modify the parameter the cursor moves on the right PARAMETER FILTER 0 5 s PRESS UNIT Pa BACK LIGHT 100 By using the UP and DOWN key choose the unit to be displayed UNIT Pa mbar mmHg Validate with the ENTER key PARAMETER FILTER 1 0 s PRESS UN...

Page 24: ... by pressing and old the ENTER during 3 seconds 500 0 Pa 000 0 The PARAMETER menu is displayed By using the UP and DOWN key select BACK LIGHT parameter PARAMETER PRESS UNIT Pa BACK LIGHT 100 AUTO OFF 05 min Then press the ENTER key to modify the parameter the cursor moves on the right PARAMETER PRESS UNIT Pa BACK LIGHT 100 AUTO OFF 05 min By using the UP and DOWN key modify the backlight value bet...

Page 25: ...n Then press the ENTER key to modify the parameter the cursor moves on the right PARAMETER PRESS UNIT Pa BACK LIGHT 100 AUTO OFF 05 min By using the UP and DOWN key to modify the switch off delay value PARAMETER PRESS UNIT Pa BACK LIGHT 100 AUTO OFF 07 min Validate with the ENTER key PARAMETER PRESS UNIT Pa BACK LIGHT 100 AUTO OFF 07 min To return to the measurement mode press CANCEL 500 0 Pa 000 ...

Page 26: ...ameters configuration menu RS232 BAUDRATE 57600 BITS COUNT 8 PARITY NONE Select and adjust each parameter BAUDRATE BITS COUNT PARITY HANDSHAKE and the EXPORT mode by using the UP and DOWN key and validate with the ENTER key RS232 BAUDRATE 19200 BITS COUNT 8 PARITY NONE To return to the measurement mode press CANCEL 500 0 Pa 000 0 The EXPORT mode allows modify the frame format this is to export the...

Page 27: ...unit Atm Bar In H2O InHg kg cm2 kPa Mbar mmH2O mmHg MPa Pa PSI Torr V Back light 0 100 V Auto off Disable 99 min V RS232 Baudrate 2400 V 4800 V 9600 V 19200 V 38400 V 57600 V Bits count 7 V 8 V Parity None V Even V Odd V Handshake No V Yes V Export No Yes Only some units are integrated according to the instrument full scale Simple menu access by press 1 second SPECIAL CYCLE Auto zero Print Sensor ...

Page 28: ... parameters menu by short press on the ENTER key 500 0 Pa 000 0 Select the AUTO ZERO function by using the UP and DOWN keys validate with the ENTER key SPECIAL CYCLE AUTO ZERO PRINT SENSOR DISPLAY If the full scale is more than 5 bars the instrument ask to disconnect every pressure source from the connector and validate with the ENTER key Switch off pressure Enter If Ok Note this cycle allows taki...

Page 29: ...s SPECIAL CYCLE AUTO ZERO PRINT SENSOR DISPLAY The frame format is different following if the EXPORT mode is validated or not Frame format sent on the USB or RS232 port with no EXPORT mode no PRESS 116 0 Pa 500 0 Pa 116 0 Frame format sent on the USB or RS232 port with EXPORT mode yes PRESS 1 497 bar 5 000 bar 1 497 In the EXPORT mode the frames contains punctuation separators and each measurement...

Page 30: ... after sales service diagnostic Enter into the parameters menu by short press on the ENTER key 500 0 Pa 000 0 Select the SENSOR DISPLAY function by using the UP and DOWN keys validate with the ENTER key SPECIAL CYCLE AUTO ZERO PRINT SENSOR DISPLAY The instrument displays the sensors points U 051974 PRESS 021509 To return to the measurement mode press CANCEL 500 0 Pa 000 0 ...

Page 31: ...tch and works as soon as it is plugged in It allows supplying the instrument who manage itself its charge It is protected against surges and short circuits via a thermal fuse Do not use any other type of fuse Moreover this supply can be plugged on all types of electrical plugs using different removable pins The instrument is not designed to work during the battery charge 1 2 SOFT CASE The soft cas...

Page 32: ...d the pressure source to control 1 5 PROTECTION COVER OPTION This cover in rubber material allows protect the CDP60 and absorb impacts option 1 6 HARD CASE OPTION Hard case to put away the CDP60 instrument and its accessories 1 7 PRESSURE GENERATOR OPTION Full scale 1000 Pa This instrument generates from the 6 bars network a pressure between 0 and 1000 Pa option ...

Page 33: ...or output to the CDP60 input 5 Turn the knob to the right to increase the pressure the pressure value is displayed by the CDP60 2 RECYCLING Do not dispose of the rechargeable Lithium Ion battery or the tool to the dustbin These components must be collected and recycled The crossed out wheeled dustbin means that within the EU the product must be taken to separate collection at the product end of li...

Page 34: ...the charger while it is inside a carrying case Connect the charger to a proper power source The voltage requirements are found on the product case and or packaging Do not use the charger if the cord becomes damaged Do not attempt to service the unit There are no serviceable parts inside Replace the unit if it is damaged or exposed to excess moisture About the Battery CAUTION This unit contains an ...

Page 35: ... press on the CANCEL key PRESSURE OUT OF RANGE Over Pressure Press C to return AUTO ZERO The applied pressure is too low Increase the pressure and press on the CANCEL key PRESSURE OUT OF RANGE Under Pressure Press C to return AUTO ZERO The LOW BAT light flashes the battery level is too low for reliable operation When this light is on the instrument will be quickly turned off automatically Charge t...

Page 36: ...Chapter 6 Error messages UM 24100D U User manual ATEQ CDP60 Page 26 36 ...

Page 37: ...onnections for update can be carried out by two ways USB or RS232 following the user preferences 1 1 USB CONNECTION 1 1 1 Windows XP installation 1 Start the installation program USB_VIRTUAL_232_XP EXE the opposite window appears 2 Click on Next 3 The Serial emulation port drivers will be installed at the end the opposite window appears 4 Click on Finish ...

Page 38: ...ort of your PC and on the CDP60 device 6 Switch on the CDP60 device 7 When the CDP60 device is detected its installation begins The opposite window appears select No not this time and click on Next 8 Select the Install the software automatically recommended option and click on Next 9 The installation begins ...

Page 39: ...29 36 10 When this window appears click on Continue Anyway 11 The drivers are installed 12 The ATEQ RS232 Emulation communication port is created this port will be used for each communication between the CDP60 device and the PC through an USB connector Click on Finish ...

Page 40: ...osite window appears 4 Cliquer sur Suivant 5 The Serial emulation port drivers will be installed at the end the opposite window appears 6 Click on Finish 7 Connect the USB cable on the PC USB port and on the CDP60 device 8 Switch on the CDP60 device the drivers will automatically installed 9 The message at the right bottom corner of the screen ATEQ RS232 Emulation appears to confirm the correct CD...

Page 41: ...d rate 19200 Bits count 8 Parity even 1 3 RS232 CABLE DIAGRAM 1 6 5 9 SubD 9 pins 8 4 1 RJ45 8 pins Pin 1 Not used Pin 1 Not used Pin 2 RXD Data reception Pin 2 Not used Pin 3 TXD Data emission Pin 3 Not used Pin 4 Not used Pin 4 Ground Pin 5 Ground Pin 5 RXD Data reception Pin 6 Not used Pin 6 TXD Data emission Pin 7 RTS request to send Pin 7 CTS clear to send Pin 8 Not used Pin 8 Not used Pin 9 ...

Page 42: ...Chapter 7 PC installation UM 24100D U User manual ATEQ CDP60 Page 32 36 ...

Page 43: ...L x P mm 152 x 83 x 36 Battery Lithium ion 12 6 V DC Autonomy Minimum 6 hours Electrical connexions Power supply concentric Jack Communication USB and RJ45 RS232 Pneumatics connexions Principal connector quick connector Staublï RBE03 female type Secondary connector 2 7 4 AVS type low pressure Display LCD 4 lines 60 mm x 32 mm Weight About 450 g Temperatures Use 0 C to 50 C Stock 10 C to 70 C ...

Page 44: ...___________________________ ________________________________________________________________________ ________________________________________________________________________ ________________________________________________________________________ ________________________________________________________________________ ________________________________________________________________________ _______...

Page 45: ...d key 7 I I key 7 Installation 6 L Lights 9 Low bat light 9 M Measurements characteristics 4 Measurements print 19 Menus structure 17 N Navigation keys 8 P Parameters adjust 12 Power supply 21 24 Presentation 5 Pressure generator 22 Pressure generator use 23 Pressure ranges 4 Pressure unit 13 Print 19 Protection cover 22 R Recycling 23 Reference 6 RJ45 6 RS232 6 16 RS232 frame 19 RS232 installatio...

Page 46: ...Index UM 24100D U User manual ATEQ CDP60 Page 36 36 ...

Page 47: ......

Page 48: ...This document is the exclusive property of ATEQ It may not be communicated reproduced or used without prior consent ...